PN MN239-C4 E | User Manual ENG - 27
NewLife Family
Recommended separation distances between portable and mobile RF
communications equipment and the NewLife Units
The NewLife is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the NewLife can help prevent electromagnetic interference by maintain-
ing a minimum distance between portable and mobile RF communications equipment (transmitters) and the
NewLife as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Classification
Type of protection against electric shock:
Class II
Protection from electric shock is
achieved by double insulation.
Protective earthing or reliance upon
installation conditions are not
required.
Degree of protection against electric shock:
Type BF Equipment providing a particular
degree of protection against electric
shock regarding
1) allowable leakage current;
2) reliability of protective earth connection
(if present).
Not intended for direct cardiac application.
Degree of protection against harmful ingress of
water:
Drip-proof equipment – IP21.
Protection against ingress of solid foreign
objects greater than 12.5 mm diameter, and
protection against vertically falling drops of
water.
Method of cleaning and infection control allowed:
Please refer to Maintenance section in the
NewLife Service Manuals.
Degree of safety of application in the presence of
flammable anesthetic gases:
Equipment not suited for such application.
Mode of operation:
Continuous duty.
d
=
1.2
√
P
d
=
1.2
√
P
d
=
1.2
√
P
Summary of Contents for NewLife Series
Page 290: ...NewLife Family 290 ELL MN239 C4 E SOC 65 NewLife...
Page 291: ...NewLife Family MN239 C4 E ELL 291 NewLife NewLife 1 1 1 A I B lpm E H 1 1 NewLife...
Page 292: ...NewLife Family 292 ELL MN239 C4 E 3 1 2 CAIRE NewLife NEWLIFE INTENSITY...
Page 293: ...NewLife Family MN239 C4 E ELL 293 I PINE SOL LESTOIL...
Page 294: ...NewLife Family 294 ELL MN239 C4 E NEWLIFE 30cm 1ft GFCI LPM 2LPM 1 8LPM 2LPM...
Page 298: ...NewLife Family 298 ELL MN239 C4 E 1 2 I Elite Intensity LPM 2LPM 1 8LPM 2LPM...
Page 299: ...NewLife Family MN239 C4 E ELL 299 NewLife NewLife NewLife 10 micron 1 1 4 2 3 4 6 5...
Page 304: ...NewLife Family 304 ELL MN239 C4 E 1 6 6 2 1 2 3 4 1 2 1 T 2 3 1 3 30 7 4 5 6 7...
Page 306: ...NewLife Family 306 ELL MN239 C4 E PINE SOL LESTOIL 1 2 3...
Page 307: ...NewLife Family MN239 C4 E ELL 307 10 LPM 30 cm 12 in NewLife...
Page 308: ...NewLife Family 308 ELL MN239 C4 E 30 5 cm 12 in I NewLife I...
Page 418: ...418 ENG PN MN239 C4 E User Manual NewLife Family NOTES...
Page 419: ...PN MN239 C4 E User Manual ENG 419 NewLife Family NOTES...