PN MN239-C4 E | User Manual ENG - 25
NewLife Family
Guidance and Manufacturer’s Declaration± Electromagnetic Immunity
The NewLife is intended for use in the electromagnetic environment specified below. The customer or the user
of the NewLife should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment ±
guidance
Electromagnetic
environment –
guidance
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power
supply lines
N/A
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
line(s)
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
NewLife Family requires continued
operation during power mains
interruptions, it is recommended
that the NewLife is powered from an
uninterruptible power supply (UPS)
or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A / m
3 A / m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE U
T
is the A.C. mains voltage prior to application of the test level.
Summary of Contents for NewLife Series
Page 290: ...NewLife Family 290 ELL MN239 C4 E SOC 65 NewLife...
Page 291: ...NewLife Family MN239 C4 E ELL 291 NewLife NewLife 1 1 1 A I B lpm E H 1 1 NewLife...
Page 292: ...NewLife Family 292 ELL MN239 C4 E 3 1 2 CAIRE NewLife NEWLIFE INTENSITY...
Page 293: ...NewLife Family MN239 C4 E ELL 293 I PINE SOL LESTOIL...
Page 294: ...NewLife Family 294 ELL MN239 C4 E NEWLIFE 30cm 1ft GFCI LPM 2LPM 1 8LPM 2LPM...
Page 298: ...NewLife Family 298 ELL MN239 C4 E 1 2 I Elite Intensity LPM 2LPM 1 8LPM 2LPM...
Page 299: ...NewLife Family MN239 C4 E ELL 299 NewLife NewLife NewLife 10 micron 1 1 4 2 3 4 6 5...
Page 304: ...NewLife Family 304 ELL MN239 C4 E 1 6 6 2 1 2 3 4 1 2 1 T 2 3 1 3 30 7 4 5 6 7...
Page 306: ...NewLife Family 306 ELL MN239 C4 E PINE SOL LESTOIL 1 2 3...
Page 307: ...NewLife Family MN239 C4 E ELL 307 10 LPM 30 cm 12 in NewLife...
Page 308: ...NewLife Family 308 ELL MN239 C4 E 30 5 cm 12 in I NewLife I...
Page 418: ...418 ENG PN MN239 C4 E User Manual NewLife Family NOTES...
Page 419: ...PN MN239 C4 E User Manual ENG 419 NewLife Family NOTES...