4.18.
Acquisition Control (0x8100; r/w)
Bit
Function
[5]
0 = Normal Mode (default): board becomes full, whenever all buffers
are full (see § 4.15)
1 = Always keep one buffer free: board becomes full, whenever
N-1buffers are full; N = nr. of blocks (see § 4.15)
[4]
0 = DOWNSAMPLE DISABLED
1 = DOWNSAMPLE ENABLED
allows to enable/disable downsampling, whose factor is set via
Downsample Factor register (see § 4.28)
[3]
0 = COUNT ACCEPTED TRIGGERS
1 = COUNT ALL TRIGGERS
allows to reject overlapping triggers (see § 3.3.4)
[2]
0 = Acquisition STOP
1 = Acquisition RUN
allows to RUN/STOP Acquisition
[1:0]
00 = REGISTER-CONTROLLED RUN MODE
01 = S-IN CONTROLLED RUN MODE
10 = S-IN GATE MODE
11 = MULTI-BOARD SYNC MODE
Bits [1:0] descritpion:
00 = REGISTER-CONTROLLED RUN MODE: multiboard synchronisation via S_IN front
panel signal
−
RUN control: start/stop via set/clear of bit[2]
−
GATE always active (Continuous Gate Mode) or Downsample Mode
−
Continuous Gate Mode can be used only if Channel gate mode (see § 4.12) is set in
Window Mode
−
Downsample Mode can be used prior DOWNSAMPLE FACTOR register (see
§ 4.28) valid setting (
≠
0)
01 = S-IN CONTROLLED RUN MODE: Multiboard synchronisation via S-IN front panel
signal
−
S-IN works both as SYNC and RUN_START command
−
GATE always active (Continuous Gate Mode) or Downsample mode:
−
Continuous Gate Mode: Gate always active; to be used only if Channel Gate
Mode (CHANNEL Configuration Register) is set to Window Mode
−
Downsample Mode: it is set via DOWNSAMPLE ENABLE and a value
≠
0 at
DOWNSAMPLE FACTOR register
10 = S-IN GATE MODE
−
Multiboard synchronisation is disabled
−
S-IN works as Gate signal set/clear of RUN/STOP bit
11 = MULTI-BOARD SYNC MODE
−
Used only for Multiboard synchronisation
4.19.
Acquisition Status (0x8104; r)
Bit
Function
[7]
PLL Status Flag (see § 2.5.1):
0 = PLL loss of lock
1 = no PLL loss of lock
NOTE: flag can be restored to 1 via read access to Status Register
(see § 4.36)
[6]
PLL Bypass mode (see § 2.5.1):
0 = No bypass mode
1 = Bypass mode