- 15 -
NOTE
1. The font color of the button is red, indicating that this stage is currently
ongoing.
2. The vibration motor does not start during the preloading. This stage is only
the preparation stage before the start of the test. The purpose is to make the
fan-shaped steel roller and the test mold cooperate in advance to avoid loud
noise due to the loading pressure at the beginning of the test.
3. In P1~P6 stage, the vibration motor is normal and starts according to the
set vibration frequency.
Loading and unloading settings interface
In the [Test] interface shown in Figure 6-2, click the "Loading and Unloading" button, and the
instrument enters the [Loading and Unloading Settings] interface, as shown in Figure 6-3. The
parameters and buttons are explained as follows.
Figure 6-3 Loading and Unloading Settings Interface
[Loading position signal]: It indicates the loading position signal of the cylinder to and from the
cylinder. The green indicates that the cylinder is not at the loading position, and the red indicates
that the cylinder is at the loading position.