- 10 -
V.
Test procedure
1. Connect the air source first, then the power source. Turn on the [emergency stop switch] to power
up the instrument (if there is no alarm sound, the instrument is operating normally).
2. The instrument automatically enters the [Startup] interface. Check whether the boxes in front of
"parameter download completed" and "the pressure correction table is download " are marked with
"√", as shown in Figure 6
-1. If yes, it means that the instrument parameter download is normal, and
you can proceed to the next step. Otherwise, follow the steps below to download parameters:
①
.
Click the "Enter the Test" button to enter the [Test] interface (as shown in Figure 6-2);
②
.
Click the "Help & Settings" button to enter the [Help & Settings] interface (as shown in Figure
6-3);
③
.
Click the "Parameter Setting" button to enter the [Parameter Setting] interface (as shown in
Figure 6-4). Set the loading pressure, vibration frequency and round-trip times
corresponding to each test stage. Click the "Data Download" button, and then click the
"Return" button to return to the [Help & Settings] interface;