- 11 -
④
.
Click the "Pressure Correction Table" button to enter the [Pressure Correction Table]
interface (as shown in Figure 6-9). Set seven groups of corresponding control voltage,
feedback voltage and actual pressure, and click the "data download" button;
WARNING!
In order to use this instrument safely and correctly, please carefully read the
precautions displayed on the [Startup] interface.
3. In the [Test] interface, click the "Loading and Unloading" button to enter the [Loading and
Unloading Settings] interface (as shown in Figure 6-3). Click the "loading position" button to make
the mold trolley move to the left side of the equipment to facilitate the loading and unloading of the
mold.
4. Turn off the power, open the protective door on the left side of the equipment, and remove the
fixed roller for the mold. Put the loaded test mold on the trolley of the roller compactor, put the test
mold fixed roller to fix the test mold, and close the protective door. For specific methods, please refer
to the "Feeding Steps" in the [Help notes] interface (as shown in Figure 6-11);
5. After the loading is completed, in the [Loading and Unloading Settings] interface, click the
"intermediate position" button to make the test trolley reach the middle position. After the test trolley
is in place, click the "preloading stage" button, and the instrument enters the preloading stage;
NOTE
In this mode, the vibrating motor does not work, but uses this stage to make
the pressurized cylinder and the round-trip cylinder enter the working state,
preparing for the formal start of the test.
6. In the [Test] interface, click any stage from "P1 stage" to "P6 stage". After selection, the word of
the selected button turns red, and the instrument can start working according to the parameters of
the selected test stage.
7. The instrument will automatically jump to the next stage after finishing the work of this stage. If
the number of round trips in the next stage is set to 0, the next stage will be automatically skipped
and the test will continue. After the instrument completes to the last test stage in sequence, it will
stop automatically. Click the button of the current stage, that is, click the button with the word in red
can stop the test in advance.
8. For example, click “P1 Stage” to start the test, the word of “P1 Stage” button will turn to red and