5.1.3
Loss of Display Contrast
If for some reason the instrument should loose its display contrast it is possible to reset it
easily using the procedure below.
Loss of contrast would exhibit no display at all when
the instrument is powered up.
Procedure for restoring loss of contrast:
1. Consider the SELECT buttons as labeled 1, 2, 3, and 4 from top to bottom.
2. Turn the instrument on and wait 20 seconds (or after High Voltage light flashes)
3. Press [MENU] key twice.
4. Press [SELECT “4”] one time.
5. Press [SELECT “1”] two times.
6. Press [ENTER].
7. Press [SELECT “1”] many times till contrast returns to normal.
8. Press [ENTER].
9. Arrow up to [SAVE SETUP] and save as [DEFAULT].
If the above procedure fails to restore the contrast try repeating again. Failing restoration of
the contrast may be indication of another instrument problem.
5.1.4
Preventive Maintenance
In addition to the annual battery replacement and calibration discussed below periodic
cleaning of dust is recommended depending on the environmental conditions.