Model 2408
RESISTOMAT
®
Page 53
2.6.4.1 Save Setup
Figure 2-22: Save Setup Test Conditions
Allows a set of test conditions to be stored in instrument memory or on the USB flash drive
for later recall. Test conditions are those that are user programmable in the Setup and I/O
menus (refer to paragraphs 2.6.2 and 2.6.3 above).
Stored setup conditions should always be backed-up on a flash drive or recorded on
paper.
NOTE
To store setups in the 2408’s internal memory, a flash drive must be plugged
into the USB port.
To store the current set of test conditions as a new set in unit memory, you need to select
[NEW] in the Save Setup menu and enter the identifying name up to 8 characters under which
these conditions will be stored (allowable characters from the keypad include 0 through 9
and minus, characters can also include A through Z when operating from remote control).
To save the setup under the name selected or to overwrite if the name already exists,
answer Yes or No (Figure 2-23).
Display Type
page
down
de
l
Setup
I/0
Utilities
Save
NEW
DEFAULT
USB
2408 - 1
2408 - 2
2408 - 3
2408 - 4