Model 2408
RESISTOMAT
®
Page 19
Manual Measurement Mode
In manual measurement mode each phase of a test cycle, charge, measure and discharge
is initiated manually at the users discretion. These test conditions can be DEFAULT (those
at power up), set using menus or the operator can recall a previously stored setup.
1. Connect the device under test to the 2408 test leads or other fixture being used.
2. Press [MENU] key to select menu display.
3. Press Right or Left Arrow key to select Setup menu.
4. Press
Up or Down Arrow key to select Mode in the sub menu.
5. Press [ENTER] to activate the selection field.
6. Press Right or Left Arrow key to select Manual and press [ENTER].
7. If the desired setup conditions are anything other than DEFAULT, set using menus
or proceed to Recalling Setups below before continuing with step 8. (Setup
measurement times are ignored in manual mode).
8. Initiate the charge phase by pressing [START] to apply high voltage to the
device. The High Voltage Indicator will remain on as long as a voltage is applied.
The measurement can be aborted at any time during the test cycle by pressing
[STOP].
9. Initiate a single measurement by pressing [START] again, The measured result
will be displayed on the LCD and will be updated each time [START] is pressed.
10. To discharge the device press [STOP].
11. Before removing the device press [STOP] again to end the discharge phase and turn
the High Voltage Indicator off.
Recalling Setups
1. To recall a set of test conditions previously stored:
Press [MENU] key to select menu display.
Press Right or Left Arrow key to select Utilities menu.
Press Up or Down Arrow key to select Recall Setup.
Press [ENTER] key to activate the recall entry field.
Press Up or Down Arrow key to select the desired setup. choices are:
- DEFAULT, if conditions are power up which can be changed by the user.
- USB, if setup is to be recalled from a flash drive.
- Setup Name stored in the instrument internal memory or USB.
Once the desired setup is located press [ENTER] key to finalize recall of the
setup conditions.
Summary of Contents for RESISTOMAT 2408
Page 4: ...RESISTOMAT Model 2408 Page ...
Page 6: ...RESISTOMAT Model 2408 Page ...
Page 10: ...RESISTOMAT Model 2408 Page 10 ...
Page 13: ...Model 2408 RESISTOMAT Page 13 Accessories ...
Page 14: ...RESISTOMAT Model 2408 Page 14 ...
Page 22: ...RESISTOMAT Model 2408 Page 22 ...
Page 30: ...RESISTOMAT Model 2408 Page 30 ...
Page 67: ...Model 2408 RESISTOMAT Page 67 ...
Page 84: ...RESISTOMAT Model 2408 Page 84 ...
Page 90: ...RESISTOMAT Model 2408 Page 90 ...