RESISTOMAT
®
Model 2408
Page 56
When there are more setups than can fit on the display the page down key is active. If there
is less than a whole page below, the display wraps around to the previous display. Conti-
nuing to page down will eventually return to the first display of setups.
2.6.4.3 Zero
The zeroing process automatically measures stray parameters and retains the data. It is
used to correct measurements, so that results represent parameters of the DUT alone
without test lead or fixture capacitance. Zeroing is recommended at the start of each work
day, or more often if leads, fixture, or test configuration to the DUT is changed. Zeroing
should also be performed anytime the test voltage is changed, which also includes recal-
ling a set of test conditions from memory with a different test voltage. It is important to
note that anytime the instrument is zeroed, it is done at the test voltage currently specified.
When [Zero] is selected in the menu screen and the [Enter] key is pressed, you are promp-
ted by instructions on the display, the first of which is shown in Figure 2-26.
Figure 2-26: Zeroing Procedure
Once the zeroing is initiated by pressing [START] the next message displayed is:
ZERO CAL IN PROGRESS
Once complete, typically a minute or slightly longer, the next message displayed is:
COMPLETE
HIT <MENU> KEY TO CONTINUE
ZERO AT 500 VOLTS
REMOVE ALL COMPONENTS
FROM THE TEST FIXTURE
HIT <START> WHEN READY
HIT <CNCL> TO EXIT
Summary of Contents for RESISTOMAT 2408
Page 4: ...RESISTOMAT Model 2408 Page ...
Page 6: ...RESISTOMAT Model 2408 Page ...
Page 10: ...RESISTOMAT Model 2408 Page 10 ...
Page 13: ...Model 2408 RESISTOMAT Page 13 Accessories ...
Page 14: ...RESISTOMAT Model 2408 Page 14 ...
Page 22: ...RESISTOMAT Model 2408 Page 22 ...
Page 30: ...RESISTOMAT Model 2408 Page 30 ...
Page 67: ...Model 2408 RESISTOMAT Page 67 ...
Page 84: ...RESISTOMAT Model 2408 Page 84 ...
Page 90: ...RESISTOMAT Model 2408 Page 90 ...