RESISTOMAT
®
Model 2408
Page 46
2.6.2.9 Stop on Pass
Accepts entry of a value between 0 and 300 which is the number of consecutive pass-
ing measurements that must occur to exit the measure loop before the measure time has
passed. This is only active when there is a measure time and limit specified. An entry of 0
disables the stop on pass function.
2.6.2.10 # to Average
Accepts entry of a value between 0 and 400 which is the number of measurements to
be averaged and displayed. If the value is 0 or 1 averaging is disabled and the display is
updated with each measurement, if the value is n (between 2 and 400) the average is dis-
played after n measurements. If the measure time has not elapsed after the first averaged
display, then a running average of measurements is made until the measure time is up. For
example, if n is 5, the first five measurements are made, averaged and then displayed. The
sixth measurement is averaged with the last four and the first ignored, this continues until
the user specified measurement time is complete.
When Stop on Pass is in effect, n measurements are made and averaged.
If the Stop on Pass requirement is met, the cycle stops with a Pass even though the
measure time has not passed.
If the measure time has passed and the Stop on Pass requirement has not been met,
the cycle stops with a Fail.
If the Stop on Pass requirement is not met, and the measure time has not passed,
additional measurements are made and averaged. The Stop on Pass and measure
time requirement are tested after each measurement until one or the other results in
an exit from the test cycle.
When the instrument is selected for Manual mode of operation the averaging works similar-
ly. When the [START] button is pressed to exit Charge and start measure, n measurements
are averaged. For each consecutive press of the [START] button one more measurement is
made, the last five are averaged and result displayed. The [STOP] button exits the Measure
mode to discharge the UUT normally.
Summary of Contents for RESISTOMAT 2408
Page 4: ...RESISTOMAT Model 2408 Page ...
Page 6: ...RESISTOMAT Model 2408 Page ...
Page 10: ...RESISTOMAT Model 2408 Page 10 ...
Page 13: ...Model 2408 RESISTOMAT Page 13 Accessories ...
Page 14: ...RESISTOMAT Model 2408 Page 14 ...
Page 22: ...RESISTOMAT Model 2408 Page 22 ...
Page 30: ...RESISTOMAT Model 2408 Page 30 ...
Page 67: ...Model 2408 RESISTOMAT Page 67 ...
Page 84: ...RESISTOMAT Model 2408 Page 84 ...
Page 90: ...RESISTOMAT Model 2408 Page 90 ...