Test protocol of
an inspection of
electro-medical
equipment
according to
DIN EN 62353:
2008-08
Page 1 of 2
Customer / med. Facility / practice:
Address:
Carried out:
!
Repeat inspection
!
Inspection prior initial operation
!
_____________________________________________
!
Inspection following repairs / servicing
Equipment type:
!
Hospital bed
"
Nursing bed
Protection classification:
!
I
"
II
Bed type:
DALI LOW ENTRY 24 Volt
Inventory Number:
Location:
Serial Number:
Grade level (IEC60601-2-52):
!
1
!
2
!
3
!
4
!
5
Manufacturer:
Burmeier GmbH & Co. KG
User-specific:
none
Testing equipment used (Type / Inventory Number):
1.
MPG Classification:
Class I
2.
I. Visual Inspection
OK
Not
OK
Defect Description
Visual inspection of the electriacl components
Stickers and type plates
Presence, legible
Transformer unit housing
Damage
Motor housing and lifting rods of motors
Correct position, Damage
Motor bolt
Check that the lock washer is properly in place
Hand switch
Damage, Foil intact
Cable for motor and hand control
Damage, Routing
Plug of harness / switching power supply
Available, Correct position
Visual inspection of the mechanical components
Stickers and type plates
Presence, legible
Safe working load / patient weight
Presence, legible
Trapeze pole / socket
Damage, Deformation
Bed frame
Damage, Deformation
Spring wood slats
Damage, Splinters
Castors
Damage
Reclining surface
Damage, Deformation
Wooden boards
Damage, Splinters
Welded seams
Split welded seams
Side rail
Damage, Splinters
Knurled screws
Tight fit
Wearing parts such as joints
Damage
Locking ring of backrest/thigh rest
Check that the ring is properly in place
II. Electric measurments according to DIN EN 62353:2008-08 (leakage current, direct)
Proceed as follow:
OK
Not
OK
Defect Description
1. Plug in transformer unit in test socket of testing device
2. Insert the plug of the 24 volt cable into the special adapter
(extra accessory; request from BURMEIER)
3. Connect the special adapter with the test probe of the test unit.
4. Start procedures with the measuring instrument
Limit Value
Value of first
measurement
Current actual
value
Result: Bed SK II (Type B)
0,1 mA
mA
mA
40