Test protocol of
an inspection of
electro-medical
equipment
according to
DIN EN 62353:
2008-08
Page 2 of 2
III. Performance check
OK
Not
OK
Defect Description
Performance check of the electrial components
Motion limit switching of the motors
Automatic cut-off
Hand switch, blocking functions
Test according to instructions for use
Drive motors
Abnormal Noise Development
Drive motors
Test according to instructions for use
Strain relief for power cable
Power cable firmly fastened
Performance check of the mechanical components
Joints and pivots
Smooth operation
Trapeze with belt strap
Secure hold
Castors
Drive and brake
Emergency lowering of the backrest
Test according to instructions for use
Side rail
Safe locking / Secure hold / Safe unlocking
Lower leg rest
Latching
Accessories (e. g. trapeze pole, trapeze)
Secure, Damage, Suitability
Result of the inspection:
All values within permissible range:
!
yes
!
no
Next inspection date:
Inspection passed:
!
yes
!
no
If inspection was not passed:
!
Defective, do not use the bed! => Repair
!
Defective, do not use the bed! => Take out of Service
!
Bed does not meet the safety standards
Test approval sticker applied:
!
yes
!
no
Part of this inspection report:
Comments:
Inspected on:
Inspector:
Signature:
Evaluated on:
Operator / Expert:
Signature: