5 | Installation
Büchi Labortechnik AG
30/78
Operation Manual Rotavapor® R-250 Pro
5.5.3
Installing a non BUCHI vacuum pump (option)
Precondition:
R
The vacuum meets the technical specifications.
See Chapter 3.5 "Technical data", page 17
u
If a vacuum pump is used, prepare the vacuum
pump. See related documentation.
u
Attach the vacuum valve to the instrument.
u
Attach the vacuum hose to the vacuum valve.
u
Connect the vacuum valve and the aeration
valve with a vacuum hose.
u
Connect the vacuum valve cable to the
instrument. See Chapter 3.2 "Configuration",
page 13
5.5.4
Installing the vacuum connection glass assembly R (option)
u
Attach the vacuum hoes to the Aeration valve.