Büchi Labortechnik AG
Help with faults | 10
Operation Manual Rotavapor® R-250 Pro
57/78
10.5
Adjusting the cooling water flow sensor
Precondition:
R
The cooling water flow sensor is installed on
the instrument. See Chapter 5.8.4 "Installing
the cooling water flow sensor (option)",
page 38
R
The instrument is switched on.
R
The display does not show an error.
u
Loosen the screw.
u
Move the sensor until the display shows an
error.
u
Secure the sensor in place with the screw
u
Switch the water supply to on.
ð
The error on the display disappear.
10.6
Adjusting the level sensor
Status LED
Explanation
ON
Liquid is detected
OFF
Liquid is not detected
Precondition:
R
The sensor is connected to the instrument. See
Chapter 5.8.5 "Installing the level sensor
(option)", page 39
u
Open the protective cap.
u
Hold the Sensor next to the liquid.
u
Adjust the adjustment screw so that the LED
light is on when the sensor is near to the liquid.
10.7
Adjusting the foam sensor
Status LED
Explanation
ON
Liquid is detected
OFF
Liquid is not detected