5 | Installation
Büchi Labortechnik AG
36/78
Operation Manual Rotavapor® R-250 Pro
5.7.2
Installing the heating bath replenishment (option)
NOTE
If oil is used as heating medium do not install the heating bath replenishment.
u
Attach the water supply to the heating bath
replenishment.
u
Secure the hose in place with a hose clamp.
5.7.3
Installing the drain connection at the heating bath
NOTE
u
If a heating medium other than clean water is used:
ð
Obey local regulations and statutory requirements regarding disposal.
u
Attach the drain hose to the drain valve.
u
Install the other end on a receiving system (e.g.
drain).
5.7.4
Setting the heating bath reaction during a power interruption
Status
Explanation
On
The heating bath stays in position.
Off
The heating bath moves down.
u
Push and hold the down button.
u
Set the On /off master switch to On.
+
u
Tap the down button to change the status
u
Tap the AERATE button.
ð
The setting is saved.
ð
The menu close.
AERATE