Figure 4-1 Bench Troubleshooting Circuit
Bench Troubleshooting
1. Establish a proper connection between the Brooks Digital Series Mass
Flow Meter or Controller (using Figure 4-1 as a reference) Switch on
the power and allow the instrument to warm-up for 45 minutes. In case
of a Controller model, adjust the Setpoint to zero. Do not connect the
device to a gas source yet. Observe the output signal and, if necessary,
perform the zero adjustment procedure (See Section 3-4 zeroing
function). If the output signal does not zero properly, please contact
Brooks Instrument.
2. Connect the instrument to a source of the same gas used for it's
original calibration. Regulate the Setpoint to 100% flow and adjust the
inlet and outlet pressures to calibration conditions. Verify that the output
signal reaches its full scale value and stabilizes at that value. Vary the
command voltage over the 1 to 100% range and verify that the output
signal follows the Setpoint If possible, connect a flow measurement
device to monitor the actual flow behavior and verify the accuracy of the
mass flow instrument. If the mass instrument performs as described
above, then it is functioning correctly and the problem may lie
elsewhere.
Table 4-2 lists possible malfunctions which may be encountered during
bench troubleshooting.
For Controller Models Only
: Apply +5 Vdc to the +15 Vdc valve override pin
(pin 12) and verify that the output exceeds 100%. Connect the valve
override pin to earth and verify that the output signal falls below 2%.