Section II. VALVE STEM MALFUNCTIONS
5-3. STICKING CONTROL VALVE.
Table 5-3. Troubleshooting Valve Stem Malfunctions
TYPE OF
PROBLEM /
MALFUNCTION
PROBABLE
CAUSE
RECOMMENDED REPAIR PROCEDURES
1. THE VALVE
STEM WILL
NOT MOVE
FREELY
WHILE
DEPRESSING
AND
RELEASING
THE OXYGEN
CONTROL
LEVER
a. The valve stem is
bent.
b. The valve stem
spring is worn.
c. The valve stem o-
ring is worn.
a. Follow the oxygen control valve disassembly /
reassembly procedures to remove and replace
the valve stem.
NOTE
Always replace the valve stem O-Ring when the valve
stem is replaced.
b. Follow the oxygen control valve disassembly /
reassembly procedures to remove and replace
the valve stem spring.
c. Follow the oxygen control valve disassembly /
reassembly procedures to remove and replace
the valve stem o-ring.
Section lII. CUTTING ROD MALFUNCTIONS
5-4. COLLET PROBLEMS.
Table 5-4. Troubleshooting Collet / Cutting Rod Malfunctions
TYPE OF
PROBLEM /
MALFUNCTION
PROBABLE
CAUSE
RECOMMENDED REPAIR PROCEDURES
1. THE
CUTTING
ROD WILL
NOT FIT
INTO THE
COLLET IN
THE TORCH
a. The wrong size collet
is in the torch.
b. The collet in the torch
is damaged.
a. Inspect the collet in the torch to determine if it is the
correct size for the cutting rod to be used. If
incorrect, remove the collet and replace with the
correct collet.
b. Remove the collet from the torch and inspect for
burrs or slag buildup. If found, remove, clean and
reinstall the collet.
OR
c. Remove the collet from the torch. Use a standard
slotted screwdriver to apply pressure in the collet
slots to open up the collet. Replace the collet in the
torch.