20
Aesculap
®
MINOP inVent trocar 30°
Sterilization for the US market
■
Aesculap advises against sterilizing the device by
flash sterilization or chemical sterilization.
■
Sterilization may be accomplished by a standard
prevacuum cycle in a steam autoclave.
To achieve a sterility assurance level of 10
-6
, Aesculap
recommends the following parameters:
*Aesculap has validated the above sterilization cycle
and has the data on file. The validation was accom-
plished in an Aesculap sterile container cleared by FDA
for the sterilization and storage of these products.
Other sterilization cycles may also be suitable, how-
ever individuals or hospitals not using the recom-
mended method are advised to validate any alternative
method using appropriate laboratory techniques. Use
an FDA cleared accessory to maintain sterility after
processing, such as a wrap, pouch, etc.
WARNING for the US market
If this device is/was used in a patient with, or sus-
pected of having Creutzfeldt-Jakob Disease (CJD),
the device cannot be reused and must be destroyed
due to the inability to reprocess or sterilize to elim-
inate the risk of crosscontamination.
Storage
►
Store sterile products in germ-proof packaging,
protected from dust, in a dry, dark, temperature-
controlled area.
Technical Service
►
For service and repairs, please contact your
national B. Braun/Aesculap agency.
Modifications carried out on medical technical equip-
ment may result in loss of guarantee/warranty rights
and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95-1602
Fax:
+49 (7461) 16-5621
E-Mail:
Or in the US:
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042
Aesculap Repair Hotline
Phone:
+1 (800) 214-3392
Fax:
+1 (314) 895-4420
Other service addresses can be obtained from the
address indicated above.
Aesculap Orga Tray/Sterile container (perforated
bottom)
Minimum cycle parameters*
Steriliza-
tion method
Temp.
Time
Minimum
drying
time
Prevacuum
270 °F—
275 °F
4 min
20 min
WARNING
Risk of injury and/or malfunc-
tion!
►
Do not modify the product.
Summary of Contents for Aescualp MINOP inVent trocar 30
Page 2: ......
Page 3: ...11 10 1 4 3 2 9 8 7 5 6...
Page 165: ...163 MINOP inVent 30...
Page 167: ...165 5 MINOP inVent 1 2 6 5 3 MINOP inVent 1 2 3 5 3 4 5 6 3 1 5 1 5 1...
Page 168: ...166 Aesculap MINOP inVent 30 5 5 6 7 6 5 1 5 1...
Page 169: ...167 7 3 5 MINOP inVent 1 8 8 5 1 5 3 1...
Page 170: ...168 Aesculap MINOP inVent 30 7 MINOP inVent 1 9 9 3 8 MINOP inVent 1 10 10 7 8 1 8 3 1...
Page 171: ...169 3 5 MINOP inVent 1 11 11 4 12 12 7 1 3 1 4...
Page 173: ...171 15 5 8 6 9 5 8 3 7 12 16 16 1 9 3 8 7...
Page 174: ...172 Aesculap MINOP inVent 30 10 11 17 17 11 2 18 18 11 10 11 2...
Page 175: ...173 10 11 19 MINOP inVent 1 19 4 20 20 11 1 9 4...
Page 176: ...174 Aesculap MINOP inVent 30 Aesculap Extranet www aesculap extra net Aesculap 6 45 C...
Page 177: ...175 VAH FDA CE pH 8 www a k i org Rote Brosch re 6...
Page 178: ...176 Aesculap MINOP inVent 30 H2O2 55 C...
Page 179: ...177 TA013543 20...
Page 180: ...178 Aesculap MINOP inVent 30 FDA CE BBraun Stabimed T C F t I 15 2 pH 9 II 1...
Page 181: ...179 I 35 15 1 5 II...
Page 183: ...181 Aesculap Aesculap STERILIT I lspray JG600 STERILIT I Tropf ler JG598...
Page 185: ...183 FH620R MINOP 30 inVent PE204A 30 MINOP RT068R MINOP inVent...