7 Preparation
MN031-543-S0 EN
TissueSeal / TissueSeal PLUS COMFORT
29
Jaw insert
1.
Check the ratchet
B1
for ease of motion.
2.
Check the jaw holder
B2
for damage and corrosion.
Jaws
1.
Check the electrode surfaces
A2
for cleanliness.
2.
Check the gap between the electrode surfaces
A2
in the closed
position for uniformity.
3.
Check the electrode surfaces
A2
for damage.
HF cable
1.
Check the connector for damage and corrosion.
2.
Visually inspect the insulation for damage.
7.6.
Packing
The
packaging
must meet the following requirements:
EN (ANSI AAMI) ISO 11607/ EN 868-2...10 (formerly EN
868/ANSI AAMI ISO 11607)
Suitable for steam sterilization (resistant to temperatures up to
137 °C, with sufficient steam permeability)
Serviced at regular intervals (sterilization container)
Pack the ligation instrument in a suitable disposable sterilization
package and/or a suitable sterilization container.
Sterilization in the transportation packaging is not allowed.
Summary of Contents for TissueSeal COMFORT
Page 1: ...DE Gebrauchsanweisung TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 2: ......
Page 35: ......
Page 37: ...EN Operating Manual TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 38: ......
Page 39: ...Legend MN031 543 S0 EN TissueSeal TissueSeal PLUS COMFORT 3...
Page 40: ...4 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 EN...
Page 73: ......
Page 75: ...FR Mode d emploi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 76: ......
Page 110: ......
Page 112: ...IT Instruzioni per l uso TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 113: ......
Page 146: ...MN031 543 S0 IT TissueSeal TissueSeal PLUS COMFORT 35...
Page 148: ...ES Manual de instrucciones TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 181: ...10 Descripci n general del sistema 34 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 ES...
Page 183: ...PL Instrukcja obs ugi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 184: ......