5 Operation
MN031-543-S0 EN
TissueSeal / TissueSeal PLUS COMFORT
19
WARNING
Risk of patient injury due to hot electrode surfaces and
vapor emission!
Maintain sufficient distance between the tips of the
instrument and sensitive tissue structures, such as the
pancreas or intestine.
Ensure that hot ligation instruments are not used for
preparation.
Do not lay the ligation instrument on the patient.
Grasping, clamping and sealing tissue
WARNING
Risk of patient injury due to inadvertent activation of the
ligation instrument!
Never use the AUTOSTART function.
Do not switch on the HF current before the active
electrodes are in contact with the tissue to be
coagulated and the ratchet is closed.
1.
Place the jaws
A
on the operating site.
2.
Place the tissue to be sealed between the electrodes of the jaws
A
.
3.
Close the jaws
A
to grasp the tissue. Do not grasp too much
tissue.
The tissue is grasped.
4.
Use the two-position ratchet of the handle
B
to adjust the
pressure on the tissue to best match the amount of grasped
tissue.
The tissue is clamped.
Summary of Contents for TissueSeal COMFORT
Page 1: ...DE Gebrauchsanweisung TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 2: ......
Page 35: ......
Page 37: ...EN Operating Manual TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 38: ......
Page 39: ...Legend MN031 543 S0 EN TissueSeal TissueSeal PLUS COMFORT 3...
Page 40: ...4 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 EN...
Page 73: ......
Page 75: ...FR Mode d emploi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 76: ......
Page 110: ......
Page 112: ...IT Instruzioni per l uso TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 113: ......
Page 146: ...MN031 543 S0 IT TissueSeal TissueSeal PLUS COMFORT 35...
Page 148: ...ES Manual de instrucciones TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 181: ...10 Descripci n general del sistema 34 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 ES...
Page 183: ...PL Instrukcja obs ugi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 184: ......