5 Operation
18
TissueSeal / TissueSeal PLUS COMFORT
MN031-543-S0 EN
WARNING
Risk of patient injuries from the combustion or
explosion of flammable liquids and gases!
Avoid contact with flammable gases and liquids, such
skin cleaners, disinfectants and anesthetic gases.
1.
Connect the HF cable
B3
to the HF device and switch on the HF
device.
2.
Set the output power of the HF device.
3.
Perform a thorough visual inspection and functional test each
time before using the ligation instrument (see section 7.5, page
26).
5.2.
During the operation
WARNING
Risk of patient injury due to incorrect device settings
and limited visibility!
Set the output power of the HF device to the value
necessary for the operation.
Use only approved programs (see section 8, page 31).
Carry out operations only with adequate visibility.
Summary of Contents for TissueSeal COMFORT
Page 1: ...DE Gebrauchsanweisung TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 2: ......
Page 35: ......
Page 37: ...EN Operating Manual TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 38: ......
Page 39: ...Legend MN031 543 S0 EN TissueSeal TissueSeal PLUS COMFORT 3...
Page 40: ...4 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 EN...
Page 73: ......
Page 75: ...FR Mode d emploi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 76: ......
Page 110: ......
Page 112: ...IT Instruzioni per l uso TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 113: ......
Page 146: ...MN031 543 S0 IT TissueSeal TissueSeal PLUS COMFORT 35...
Page 148: ...ES Manual de instrucciones TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 181: ...10 Descripci n general del sistema 34 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 ES...
Page 183: ...PL Instrukcja obs ugi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Page 184: ......