7 Preparation
44
NightKNIFE / Ligator
MN031-447-S0-20101217-EN
x
the HF cable is located in a sieve tray with a
cover.
2. For NightKNIFE with exchangeable blade: remove a
new blade with insertion aid from its package and place
the blade and insertion aid in a sieve tray with a cover.
3. Restart the program.
4. After the program is finished, remove the instrument
components from the CDM.
NOTE
Risk of handle damage due to compressed air!
X
Restrict compressed air pressure to 3 bar or less for
handle drying.
5. Use filtered compressed air to dry the instrument
components.
7.6. Inspection
These products are subject to wear when used as intended,
depending on the intensity of use. This wear arises from the
design and construction of the instruments and is
unavoidable.
Replace the product if it shows externally visible defects or
does not function as described in this manual. Please advise
the manufacturer or the manufacturer’s authorized
representative in such cases.
X
After cleaning, visually inspect each of the instruments
and test them for proper operation.
X
Replace any damaged parts.
MN031-447-S1 EN