5 Operation
24
NightKNIFE / Ligator
MN031-447-S0-20101217-EN
Grasping, clamping and sealing tissue
WARNING
Risk of patient injury due to inadvertent activation of the
ligation instrument!
X
Never use the AUTOSTART function.
X
Do not switch on the HF current before the active
electrodes are in contact with the tissue to be
coagulated.
1. Place the jaw insert
A
on the operation site.
The jaw insert
A
can be rotated and locked in steps of 45°.
2. Turn the fluted knob
B2
to set the angle of the jaw
insert
A
.
3. Place the tissue to be sealed between the electrodes of
the jaws
A1
.
4. Close the jaws
A1
to grasp the tissue.
ª
The tissue is grasped.
5. Use the three-position ratchet of the handle
C
to adjust
the pressure on the tissue to best match the amount of
grasped tissue.
ª
The tissue is clamped.
MN031-447-S1 EN