7 Preparation
40
NightKNIFE / Ligator
MN031-447-S0-20101217-EN
Handle
NOTE
Risk of material damage to the handle from metal
brushes!
X
Use only the plastic brushes supplied with the
instrument to clean the shank tube.
X
If necessary, use a steam sprayer for cleaning.
6
3
2
1
4
1. Actuate the handle until it latches, and use the cleaning
brush to clean the guide rails
6
.
2. Using the cleaning brush, clean all of the marked
positions on the handle from the inside.
MN031-447-S1 EN
6
3
1
4
X
X
1. Actuate the handle until it latches, and use the cleaning
brush to clean the guide rails
6
.
2. Using the cleaning brush, clean all of the marked
positions on the handle from the inside.