MN031-447-S0-20101217-EN
4. Assembly
WARNING
Risk of patient injury from non-sterile ligation
instruments!
X
The ligation instrument is not sterile as delivered. Clean
and sterilize the instrument before using it.
Pos: 6 /679-BOWA/HF-Instrumente/Ligator "NightKNIFE"/4.1 Montage der Klinge @ 5\mod_1277985392590_6.doc @ 40655 @ 2
4.1.
Attaching the replaceable blade with the
insertion aid (only for NightKNIFE with
replaceable blade)
;
The blade and insertion aid have been cleaned and
disinfected ( see section 7.5, page 42).
1. Remove the blade and insertion aid
E
from the sterile
package.
2. Rotate the jaw insert
A
with the fixed jaw
A1
facing
upward to open the jaws.
3. Insert the fixed jaw
A1
between the blade
E2
and the
insertion aid
E1
and press the insertion aid
E1
downward in the arrow direction to release the blade.
MN031-447-S1 EN