2.2.1. HF
device
X
Use only approved HF devices and programs (see
section 8, page 51).
X
Observe the operating instructions of the HF device and
the general instructions for electrosurgical operations.
Improper use of HF current can lead to endogenous and
exogenous burns and explosions:
X
Carry out electrosurgical operations only with insufflation
using a non-flammable gas (CO
2
).
X
Avoid direct skin contact with HF cables.
X
Avoid contact with flammable gases and liquids.
2.2.2. HF
cable
Improper use of HF cables can lead to patient injuries:
X
Never lay the HF cable on the patient’s skin.
X
Connect the ligation instrument for coagulation before
switching on the HF generator.
X
When plugging or unplugging the HF cable, always
grasp the connector directly.
X
Use only HF cables that are in perfect condition. Never
use a defective HF cable.
The HF cable may cause interference to imagery on
monitors.
X
Never route the HF cable alongside a camera cable.
X
Do not lay the HF cable in loops.