4 Assembly
MN031-447-S0-20101217-EN
NightKNIFE / Ligator
21
4.2.
Assembling the ligation instrument
1. Screw the jaw insert
A
fully into the shank tube
B
.
2. Check that there is no gap between the jaw insert
A
and
the shank tube
B
.
3. Close the jaws
A1
with your fingers and insert the shank
tube
B
in the handle
C
:
x
With NightKNIFE: ensure that the alignment
spring
B4
and the alignment groove
C5
line
up with each other.
x
Let the shank tube
B
click into the handle
C
.
4. Check the ligation instrument for proper gripping
operation by actuating the handle
C
.
5. Using the ratchet lever
C1
, open the jaws
A1
.
6. With NightKNIFE: check the initial position and ease of
motion of the blade by actuating the trigger
B3
.
7. Check that the fluted knob
B2
can be turned.
8. Connect the HF cable
D
to the handle
C
and check that
it is securely attached.
x
When plugging or unplugging the HF cable,
always grasp the connector directly.