This message indicates a normal condition. It is displayed
when RF power is requested for a non-OI catheter but
the temperature setting has not been changed from the
default.
• Adjust temperature setting to an appropriate value
before requesting RF delivery.
Communication has been interrupted between the
Maestro and Pump while the Maestro is in
STANDBY
mode. When the Maestro 4000 and MetriQ are con-
nected and have an OI catheter attached, they oper-
ate in Automatic mode. In this state the pump will
display “Automatic” at the top of the display screen.
If they become disconnected, the system will switch
to manual mode and the pump will display “Manual”.
In manual mode the Maestro 4000 and MetriQ Pump
operate independently. The Maestro 4000 will not
wait for the pump to start ablation flow before deliver-
ing RF Power to the catheter.
If the message is displayed while the communications
cable is connected, do the following:
1. Check the cable connections between the
Maestro 4000 and MetriQ Pump. Tighten the
thumb screws on the connectors.
2. Replace the communications cable.
The default Power/Temperature mode for non-OI cath-
eters is Temperature mode. If Power mode is selected
instead, the Maestro reduces power by 50% and dis-
plays this message when the temperature exceeds the
temperature set point.
• Verify that power-control mode is the desired
mode of operation.
• Decrease the power setting and/or increase the
temperature setting as appropriate.
M09-SET TEMPERATURE
M11-RF REDUCED 50%, MAX TEMP REACHED
M10-RECONNECT PUMP
Cause:
Action:
Cause:
Action:
Cause:
Action:
All cited trademarks are the property of their respective owners. CAUTION: The law restricts these devices to sales by or on the
order of a physician. Indications, contraindications, warnings and instructions for use can be found in the product labeling
supplied with each device. Information for the use only in countries with applicable health authority product registrations.
EP-252604-AA JUN2014 Printed in Germany by medicalvision.
www.bostonscientific.com
© 2014 Boston Scientific Corporation
or its affiliates. All rights reserved.
DINEP2211EA
40
This message indicates a normal condition. It is displayed
when RF power is requested for a non-OI catheter but
the temperature setting has not been changed from the
default.
• Adjust temperature setting to an appropriate value
before requesting RF delivery.
Communication has been interrupted between the
Maestro and Pump while the Maestro is in
STANDBY
mode. When the Maestro 4000 and MetriQ are con-
nected and have an OI catheter attached, they oper-
ate in Automatic mode. In this state the pump will
display “Automatic” at the top of the display screen.
If they become disconnected, the system will switch
to manual mode and the pump will display “Manual”.
In manual mode the Maestro 4000 and MetriQ Pump
operate independently. The Maestro 4000 will not
wait for the pump to start ablation flow before deliver-
ing RF Power to the catheter.
If the message is displayed while the communications
cable is connected, do the following:
1. Check the cable connections between the
Maestro 4000 and MetriQ Pump. Tighten the
thumb screws on the connectors.
2. Replace the communications cable.
The default Power/Temperature mode for non-OI cath-
eters is Temperature mode. If Power mode is selected
instead, the Maestro reduces power by 50% and dis-
plays this message when the temperature exceeds the
temperature set point.
• Verify that power-control mode is the desired
mode of operation.
• Decrease the power setting and/or increase the
temperature setting as appropriate.
M09-SET TEMPERATURE
M11-RF REDUCED 50%, MAX TEMP REACHED
M10-RECONNECT PUMP
Cause:
Action:
Cause:
Action:
Cause:
Action:
All cited trademarks are the property of their respective owners. CAUTION: The law restricts these devices to sales by or on the
order of a physician. Indications, contraindications, warnings and instructions for use can be found in the product labeling
supplied with each device. Information for the use only in countries with applicable health authority product registrations.
EP-252604-AA JUN2014 Printed in Germany by medicalvision.
www.bostonscientific.com
© 2014 Boston Scientific Corporation
or its affiliates. All rights reserved.
DINEP2211EA
40
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
39