Tubing is leaking
• Ensure catheter to tubing kit Luer fitting is correctly connected
• Replace tubing set for all other leaks
There is fluid in burette, but no return fluid is seen dripping back into burette chamber
(you may see an impedance rise)
• Check that the tubing is not kinked or occluded
• Check that the catheter shaft is not kinked
Blue float seal is stuck on bottom of burette chamber
• Turn pump off first
• Open the occlusion bed latch slightly, this should release the float seal
• Check for leaks
• Ensure there is ~10 cc of fluid in burette chamber when pump is on
“Knocking” sound
(Note: Pump makes a continuous audible churn, this is normal)
• Ensure tubing set is pulled taut inside the occlusion bed
• If knocking continues, change the tubing set
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
CircuCool
™
Fluid Circulation Pump
Trouble-Shooting Guide
30