Display
Description
Action
D06-Sensor Failure
Input temperature <20°C +/- 2°C
Replace Catheter.
Replace Catheter
or >120°C +/- 5°C
(bad temp sensor or reference).
D07-Temp Out Of Range
Attempt to enter RF Power On mode
Ensure correct Catheter
with a measured temperature outside the positioning. Check
rangeof 19.5°C to 43.5°C or less than 30
measured temperature
seconds have elapsed since last ablation
in
READY
mode.
and the temperature is above the set point.
D08-Check Pod
Missing or bad pod connection.
Check connections.
Restart system. Call Boston
Scientific if error recurs.
D09-Replace Catheter
Invalid catheter ID.
Replace Catheter.
Invalid ID
D10-Check Remote
There is a communications problem
Restart system. Check
Control Panel
between the Remote Control Unit
connections. If error recurs,
and the Controller.
disconnect Remote Control
Unit and contact
Boston Scientific.
M01-Limited Output
TEMPERATURE MODE:
Maximum
Check DIP electrode
power is not sufficient to achieve the
placement and
desired temperature. However ablation
application prior to
will continue.
continued delivery.
POWER MODE:
Maximum
temperature
is reached. RF power will be titrated
down by the generator to avoid
excessive temperature.
HIGH Impedance
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
Maestro 3000
™
Cardiac Ablation System
Trouble-Shooting Guide