7)
Install one DIP electrode (or two,
if using a high power catheter) into
either (or both) of the “Indifferent
Electrode” receptacles on the front
of the Pod.
8)
Receive the Catheter Cable from
the sterile field.
9)
Connect Catheter Cable to the
front of the Pod.
10) Optional:
If the Remote Control
Unit is used, connect a Hospital
Grade power cord from the Remote
Control Unit to a properly grounded
AC wall outlet designated “Hospital
Grade” or “Hospital Only.”
11) Optional:
If the optional Remote
Control Unit is used, connect the
Remote Control cable to the back
panel of the Controller.
12) Optional:
If the Remote
Control Unit is used, connect
the cable from the Controller
to the rear panel of the Remote
Control Unit.
13)
Turn on the Controller by pressing
the power switch on the front panel of
the Controller to the “I” position.
Maestro 3000™
Cardiac Ablation System
Maestro 3000™ Controller
Maestro 3000™ Pod
Maestro 3000™ Remote Control Unit
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
Maestro 3000
™
Cardiac Ablation System
Set Up Guide
22