Step 3. Connecting the Generator to the Recording System with the M004 653S 0 Cable
Step 4. Connecting the Generator to the Filter Module
Step 1. Connecting the Catheter to the Filter Module and Pod with the M00 4620 0 Cable
a. Connect the
black
620
cable plug into the IntellaTip MiFi XP Ablation Catheter
b. Connect the
red
620
cable with the
115
cable
c. Connect the
red
115
cable to the front port marked “Catheter” on the generator
d. Connect the yellow
620
cable plug into the port marked “Catheter Cable” on the Filter Module
Step 2. Connecting the Filter Module to the Recording System with the M004 653S 0 Cable
a. Connect the M004
653S
0 Cable plug into the green port marked “Recording System”
on the Filter Module. Connect the second M004
653S
0 Cable to Pod
b. Connect the 3 Mini Electrodes (ME’s) via the three pins D, 2, and 3 to the Catheter Pin
Connection Box. Pin 4 is not used.
a. Connect the M004
653S
0 Cable plug to the Generator
b. Connect the conventional bipoles via the pins 1, 2, 3 and 4 to the Catheter Pin Connection Box
a. Connect the shrouded pin-end (2mm shrouded pin) of Reference Cable to the port marked “Reference Cable” on the front
of the Filter Module
b. Connect the 2 prong plug end of the Reference Cable to the Dispersive Indifferent Patch (DIP)
c. Connect the 2 prong plug end of the Reference Cable to
39D82X
Reference cable (required with Valley Lab DIP)
d. Connect the
39D82X
Reference cable to the front of the Generator with ‘Blue’ plug which has the DIP symbol
CATHETER PIN CONNECTION BOX
Cable M004 653S 0
Global Port
Digital ECG
Connection Cable
Stockert
Global Port
Serial Computer
Interface Cable D-type
EGM Connection
Cable
ValleyLab
Ground Pad
M004 354 0
Stockert EP-Shuttle
Radiofrequency Generator
IntellaTip MiFi™ XP Cable
M004 620 0
IntellaTip MiFi™ XP Cable
M004 115 0
IntellaTip MiFi™ XP
Ablation Catheter
Cable: Filter Module to Recorder
M004 653S 0
Reference Cable
M004 3636 0
Reference Cable
M004 39D82X 0
IntellaTip MiFi™ XP Filter Module
M004 1212 0
Cable: M004 653S 0
Step 2
(Red)
(Yellow)
(Green)
Reference Cable:
Filter Module to RF Pod
M004 3636 0
Step 4
IntellaTip MiFi XP Cable
M004 620 0
Step 1
Cable:
M004 653S 0
Step 3
In order to connect 2 ValleyLab
®
Ground Pads, adaptor EO5O7B (provided by Covidien) is needed.
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
19
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
18