![Boston Scientific Maestro 3000 Connection Manual Download Page 28](http://html.mh-extra.com/html/boston-scientific/maestro-3000/maestro-3000_connection-manual_2796091028.webp)
1)
Connect the Model 2071 TTT Box
to the Model 822T APM™ Device.
2)
Turn the knob on the front of the
CircuCool Pump counterclockwise
until loose and open the front of
the occlusion bed.
3)
Lift the latch handle and open
the occlusion bed.
4)
Using sterile technique, hold the
catheter lumens on the bottom and
connect and tighten the luers to
the catheter luers.
5)
Using sterile technique, connect
the sterile cable to the catheter.
6)
Hand off the tubing set
to someone in a non-sterile field.
7)
Hand off the catheter cable to
someone in a non-sterile field.
8)
Connect catheter cable to
the TTT Box (model 2071).
Turn on the TTT Box;
verify “on” light is illuminated.
9)
Insert the hard Peristaltic
tubing into the lower portion
of the occlusion bed.
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
CircuCool
™
Fluid Circulation Pump
Set Up Guide
27