Maestro 4000
™
Cardiac Ablation System
Controller Setup Guide
Note: This Guide is not intended to replace the Maestro 4000 Operator’s Manual. Thoroughly review the manual before
installing or operating this system. Complete installation diagrams are shown in the manual. Not all components shown in
this setup guide will be used with every system.
Maestro 4000 Controller Front Panel
Maestro 4000 Controller Back Panel
Power Cord
Pod
Cable
On/Off
Switch
Pump Remote
Volume
Foot Switch
Recording System
1)
Connect EP Recording sys-
tem cable to the back panel
of the Controller.
5)
Optional: If a Remote Con-
trol Unit is used, connect the
Remote Control cable to the
Remote Connector.
2)
Connect a Hospital Grade
power cord from the rear
panel of the Controller to
properly grounded AC wall
outlet designated “Hospital
Grade” or “Hospital Only”.
3)
Optional: If the optional
footswitch will be used, con-
nect it to the Foot Switch
Cable Connector.
4)
Optional:
If a MetriQ
™
Pump is used, connect the
Pump cable to the Peristal-
tic Pump Connector.
7)
Connect the cable from
the Controller to the rear of
the Remote Control Unit.
6)
Optional: If a Remote Con-
trol Unit is used, connect a
Hospital Grade power cord
from the Remote Control Unit
to a properly grounded AC
wall outlet designated “Hospi-
tal Grade” or “Hospital Only”.
8)
Connect the recording
system cable (model 653S)
to the back of the Pod.
33
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
Maestro 4000
™
Cardiac Ablation System
Controller Setup Guide
32