Alphanumeric
Display Window
Maestro 3000™ System
Diagnostic Codes
The Maestro 3000™ Cardiac Ablation System will display diagnostic codes during
STANDBY
and
DELIVER
modes of operation. The code will appear in the Alphanumeric
Display Window, located on the front panel of the System as shown above. Below is a
description of the various diagnostic codes and the action to be taken to clear the code.
These and complete operating instructions may be found in the Operators Manual.
Display
Description
Action
D01-Low Impedance
Displayed impedance below the low
Check all connections.
impedance threshold of 50
Ω
for a standard catheter and 25
Ω
when using a high power catheter.
D02-High Impedance
Displayed impedance above the
Check application of
impedance setting (150
Ω
– 300
Ω
).
DIP electrode. Check all
connections. Remove
and inspect catheter.
D03-Limit Exceeded
V > 1.1 x Vmax or I > 1.1 x Imax
Restart System.
or P > 1.2 x Pmax.
Contact Boston
Scientific if error
recurs.
D04-Temp Above Set Point
Displayed temperature is greater than
Check set temperature.
set point by at least 5°C
for more than four seconds.
D05-Excessive Temperature
Displayed temperature greater than 95°C Check catheter.
for more than one second or measured
Reduce power and/or
temperature exceeds setting in
set temperature.
Power Control
mode.
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
Maestro 3000
™
Cardiac Ablation System
Trouble-Shooting Guide