OCCLUSION SENSOR TRIGGERS, BUT SALINE FLOWS FREELY FROM THE TIP.
• Open the pump door completely.
• Pull the tubing out of the bubble detectors on either side of the pump head. Gently stretch the cable and reinsert into
the bubble detectors. LEDs for both detectors should be green and tubing should be straight where it passes over the
occlusion sensor.
PUMP REMAINING VOLUME LEVEL DOESN’T MATCH AMOUNT IN SALINE BAG
• Open pump menu and check that the selected bag size matches the actual size.
• If “Start New Case” was selected at startup, but bag was not replaced after previous session; or bag was replaced
and “Continue Previous Case” was selected, the volume remaining will be incorrect.
• “New Saline Bag” was selected from the menu, but saline bag was not changed.
PRE-RF OR POST-RF TIME DIFFERENT THAN EXPECTED
• Open pump menu and check the POST-RF and PRE-RF settings.
PUMP ALARM IS INAUDIBLE OR TOO LOUD
• Open pump menu and increase or decrease the volume level. The level will be saved for future sessions.
CATHETER NOT RECOGNIZED. (OI CATHETER IS ATTACHED, BUT MAESTRO SAYS “FINDING CATHETER…”)
• Check connections at both ends of catheter cable.
• Check cable number printed on the catheter cable. If it does not read “671” replace the cable with a 671 cable.
• Replace existing 671 cable with a new 671 cable.
• Replace OI catheter.
• Replace Pod.
RED LIGHT IN THE BUBBLE DETECTOR DOES NOT TURN FROM RED TO GREEN
• Gently stretch the tubing and push into slot on bubble detector.
• Close and latch door to push tubing into place. Open door again to check that LEDs have turned green.
• Check the tubing for blemishes where it passes through bubble detectors. Replace tubing if it has blemishes.
PUMP IS UNRESPONSIVE OR ERROR MESSAGE ON PUMP CANNOT BE CLEARED
• If the Maestro is displaying an error, clear the error on the Maestro, then clear the pump error.
• This may occur if pump is powered on with the catheter in the patient. Remove the catheter from patient,
clear the error, enable standby flow and re-insert catheter.
• If the above steps don’t clear the error use the power switch on the back of the pump to
turn the pump off and back on. If a procedure is in process select “Continue Previous
Case “ from the start-up menu.
• Resume procedure.
All cited trademarks are the property of their respective owners. CAUTION: The law restricts these devices to sales by or on the
order of a physician. Indications, contraindications, warnings and instructions for use can be found in the product labeling
supplied with each device. Information for the use only in countries with applicable health authority product registrations.
EP-252602-AA JUN2014 Printed in Germany by medicalvision.
www.bostonscientific.com
© 2014 Boston Scientific Corporation
or its affiliates. All rights reserved.
DINEP2209EA
36
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
35