10)
Slide the hard Peristaltic tubing
into the upper slot of the occlusion
bed. Make sure the Peristaltic
tubing is inside the occlusion bed.
11)
Close the occlusion bed latch.
Make sure that the burette chamber
is not distorted by the holder.
12)
Close the door. Push the knob
in and turn it clockwise until snug.
Ensure the burette chamber and
pulse dampener are in their slots.
13)
Using a syringe, place 30cc of
sterile normal saline 0.9% or D5W,
into the burette chamber.
14)
Check that the blue float seal is
floating, then turn on the pump.
15)
As the system circulates,
the fluid level will drop. When it
drops by 10cc, add an additional
10cc of sterile normal saline 0.9%
or D5W.
The unit is ready to use.
16)
Caution: Perform system integrity check.
Always prime the
tubing prior to inserting the catheter into the patient.
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
CircuCool
™
Fluid Circulation Pump
Set Up Guide
28