Guidance and manufacturer‘s declaration
Electromagnetic emission
The
medizinisches Bett
is intended for use in the electromagnetic environment specified below.
The customer or the user of the
medizinisches Bett
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11 (partly)
Group 1
The medical used bed uses RF energy only for its internal function. Therefore, its RF emissi-
ons are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11 (partly)
Class B
The
medizinisches Bett
is suitable for use in all establishments other than domestic and
those directly connected to the public-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Electromagnetic immunity
The
medizinisches Bett
is intended for use in the electromagnetic environment specified below.
The customer or the user of the
medizinisches Bett
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrostatic transient/
burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply ines
± 1 kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
< 5 % U
T
(>95 % dip in U
T
) for 0,5 cycle
40 % U
T
(60 % dip in U
T
) for 5 cycles
70 % U
T
(30 % dip in U
T
) for 25 cycles
< 5 % UT
(>95 % dip in U
T
) for 5 sec
< 5 % U
T
(>95 % dip in U
T
) for 0,5 cycle
40 % U
T
(60 % dip in U
T
) for 5 cycles
70 % U
T
(30 % dip in U
T
) for 25 cycles
< 5 % U
T
(>95 % dip in U
T
) for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the medizinisches Bett requires continued
operation during power mains interruptions, it is
recommended that the medizinisches Bett be
powered from an uninterruptible power supply
or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: U
T
is the a. c. mains voltage prior to application of the test level.
Electromagnetic immunity