39
tems, oxygen concentrators,etc.). In this case, it would
be necessary, to deactivate all bed functions by means
of the integrated locking device, until the treatment is
completed.
Important:
The beds are not equipped with any particular con-
nections that provide for a potential equalization. Electrical medi-
cal devices connected to the patient´s intravascular or intracar-
diac system must not be used. The operator of the medical
device shall be responsible for the conformity of the combinati-
on of the devices with the requirements of DIN EN 60601-1-1.
Particular features:
The advanced lifting technology installed under the lying surface
makes these beds stand out with an elegant design. This design
includes the whole spectrum of technical comfort. The adjust-
ment functions of the lying surface allow for virtually every posi-
tion – even the sitting position is no problem.
The model equipped with a central braking is also available
without covered castors. The object beds of the practico series
are as well available with splitted as continuous wooden side
rails, end panels for the head- and foot end. The lying surface
of the model practico is available with a 4 section lying surface.
The electrical adjustment of the back and leg rest can be easily
made by means of an automatic triple function.
The health bed practico is optionally available including a
Trendelenburg function or a comfort seating position.
Assembly of the bed
Please remove all packaging from the health care bed before
carrying out the assembly.
1. Remove all connections toward the transport packaging
2. Position the base with its locked castors and bring it into
the uppest position.
3. When mounting the unpartitioned lying surface, just
insert the ball bearings positioned at the top section of
the scissors into the provided u-channel at the bottom
section of the lying surface. Now remove the spring cot-
ter pin from the base frame. Use the bolt to adjust the
lying surface between the two brackets. Use the spring
cotter pin to protect the bolts against unfastening.
4. The net cable has to be fastened to the lying surface with
the provided pull relief.
5. Connect the mains plug and the actuator with the main
motor and bring the lying surface to its uppest position.
> Bock Top Advice
Bock supports you with
a checklist in accordance
with DIN EN 62353 (inclu-
ded to this user´s manu-
al). This is time saving and
gives you the needed cer-
tantiy for a thorough carry
out of the maintenance.