BT-780 Operation manual
45
P/N : 780-ENG-OPM-EXP-R00
Bistos Co., Ltd.
2020.03
Source
Alarm message
P
Causes and countermeasures
appropriate for the patient.
PR Too High
PR value is higher than the upper alarm limit or lower than the
lower alarm limit. Check the patient's physiological condition, and
check if the patient category and alarm limit settings are
appropriate for the patient.
PR Too Low
NIBP
NIBP signal weak
M*
NIBP value is higher than the upper alarm limit or lower than the
lower alarm limit. Check the patient's physiological condition, and
check if the patient category and alarm limit settings are
appropriate for the patient.
NIBP-Sys Too High
NIBP-Sys Too Low
NIBP-Mean Too High
NIBP-Mean Too Low
NIBP-Dia Too High
19.2 Technical alarms
Source
Alarm message
P
Causes and countermeasures
System
Battery Low
H
Connect to AC power supply, and charge the battery, and power
with the battery as needed after fully charged.
ECG
ECG Comm. Stop
H
ECG module failure, or communication failure between the
module and the host; please restart the device.
ECG Comm. Error
H
ECG Config Error
H
ECG Selfcheck Error
H
ECG Lead Off
M*
The electrodes are not connected to the patient firmly or fall off,
or lead wires and the main cable fall off. Check the connection of
electrodes and lead wires.
ECG YY OFF (YY is a lead
name)
M*
Temp
TEMP1 Sensor Off
L
The temperature sensor falls off from the patient. Check the
sensor connection.
TEMP2 Sensor Off
L
SpO
2
SpO
2
Comm. Stop
H
SpO
2
module failure, or communication failure between the
module and the host; please restart the device.
SpO
2
Comm. Error
H
SpO
2
No Sensor
L
SpO
2
sensor falls off from the patient or monitor, malfunctions,
or sensor other than specified in this Manual is used. Check the
sensor mounting position, whether the sensor is damaged or
sensor type. Reconnect the sensor or use new sensor.
SpO
2
Sensor Off
L
SpO
2
Search Timeout
L
SpO
2
Search Pulse
L
Sensor signal is poor or too weak. Check the patient's condition,
and place the sensor in a suitable position. If the failure persists,
replace the sensor.
NIBP
NIBP Comm. Stop
H
NIBP module failure, or communication failure between the
module and the host; please restart the device.
NIBP Comm. Error
H
NIBP Selfcheck error
H
NIBP CFG Error
H
NIBP system error
H
If failure occurs during measurement, the system can’t analyze
and calculate. Check the patient's condition, check the
connections or replace the cuff, and then re-test.
Measurement timeout
L
Cuff type error
L
The used cuff does not match the set patient category. Verify the
patient category and replace the cuff.
NIBP
Cuff loose or no cuff
L
NIBP cuff isn’t placed or connected properly, or there is gas leak.
Cuff leak
L
Check cuff and inflation tube.
Air pressure error
L
Ambient atmospheric pressure is abnormal. Confirm that the
environment complies with the monitor's specifications, and
check whether there are special reasons affecting ambient
pressure.
NIBP over range
L
The measured blood pressure of the patient exceeds the
measuring range.
NIBP signal weak
L
Patient’s pulse may be weak or cuff is too loose. Check the
condition of the patient, and place the cuff in a suitable position.
If the failure persists, replace the cuff.
NIBP signal unstable
L
Excessive movement may result in too much motion artifact or
interference in the signal during measurement.
NIBP signal saturated
L
Motion signal amplitude is too large due to movement and other
reasons.
NIBP over pressure
L
Cuff overpressure, and gas blockage may occur; check the gas
path, and then re-measure.
Module reset failed
L
NIBP module reset error; check the gas path is blocked, and then