BT-780 Operation manual
29
P/N : 780-ENG-OPM-EXP-R00
Bistos Co., Ltd.
2020.03
Figure 8-4:
“
RESP Setup” menu
8.5.1 Setting apnea time
Apnea alarm is a high level alarm for monitoring the apnea. In “RESP Setup” menu, set “Apnea Delay” to an appropriate value and
set the apnea alarm time. When the apnea time of the patient is longer than the set time, the monitor will trigger an alarm. Set
time can be set“20s” / “25s” / “30s” / “35s” / “40s” /“45s” /“50s” /“55s” /“60s”, default apnea alarm time is 20s.
8.5.2 Adjusting wave gain
In “RESP Setup” menu, select “Wave Gain”, and set the wave gain: the greater gain, the higher wave amplitude. Gain can be set
“x0.5” / “x1” / “x2”,the default is “x1”.
8.5.3 Setting sweep speed
In “RESP Setup” menu, select “Wave speed”, and set the sweep speed: the faster sweep speed, the smoother wave. The wave
speed is “6.25mm/s” / “12.5mm/s” / “25mm/s”. The default is “12.5mm/s”.
8.5.4 Setting calculated channel
In “RESP Setup” menu, select “Calculated Channel”, and set the calculated channel. The calculated channels are “RA-LA”, “RA-LL”,
“LA-RL” and “LL-RL”. The default is “RA-LA”.
8.5.5 Setting sensitivity
In “RESP Setup” menu, select “sensitivity”, and set the sensitivity. The sensitivity is “1” / “2” / “3” / “4” / “5”. The default is “2”.
8.6 Alarm setup
Select “Alarm Setup >>
”→
“
Alarm Setup
”
interface to set respiration related alarms; see 6.5 Alarm Setup for the setting method.
9 PR
9.1 Overview
The mechanical activity of the heart causes arterial pulsation, and PR (pulse rate) value can be obtained by measuring this pulsation.
PR value can be obtained through SpO
2
measurement.
The average calculation of the heart rate is the direct average method. The refresh rate is every 1 second.
9.2 Display
The color of PR parameter area is same as SpO
2
parameter color of PR source, as shown in Fig. 9-1:
Figure 9-1: PR parameter display
9.3 Setting PR sound
Select SpO
2
parameter area or SpO
2
wave area
→
“
SpO
2
Setup” menu;
Select “Pulse Volume” to set “Pulse Volume” to 0~9. Select 0 to turn off the pulse volume, and select 9 to set the maximum volume.
NOTE
HR sound has higher priority than PR sound. When HR makes a sound, PR won’t. When HR sound set to 0, PR can
make a sound.
9.4 Alarm setup
Select PR parameter area
→
“SpO
2
Setup
”
menu
→
“
Alarm Setup >>” to enter the “Alarm Setup” interface, and set PR alarm
switch, alarm level and upper/lower alarm limit. See 6.5 Alarm Setup for detailed setting method.