BT-780 Operation manual
39
P/N : 780-ENG-OPM-EXP-R00
Bistos Co., Ltd.
2020.03
Environment and power meet the requirements.
Device and accessories have no mechanical damage.
The power supply has no wear, and the insulation is good.
Specified accessories are used.
Alarm system is functioning correctly.
Battery performance meets the requirements.
Monitoring functions are in good working condition.
Ground impedance and leakage current meet the requirements.
If the instrument function has any sign of damage, do not use this monitor for any patient monitoring. Please contact the hospital's
professional maintenance personnel or our customer service personnel.
All checks that require disassembling the instrument must be performed by qualified service personnel. Safety and maintenance
checks may also be carried out by the Company's personnel.
16.2 Viewing software version information
You can view the software version through the following steps:
Select
【
Settings
】
Smart Hotkey
→
“Settings” Menu;
Select “Monitor Info>>
”→
“Monitor Info” menu;
“Monitor Info” menu displays the software version information of the monitor.
16.3 Maintenance plan
The following tasks can only be done by qualified service personnel of Bistos. When the following maintenance is needed, please
contact your service representative. Before testing or maintenance, clean and disinfect the device.
Inspection / Maintenance Item
Frequency
Check the safety according to IEC 60601-1
At least once every two years, after replacing the power supply or
the monitor falls down.
Check all monitoring or measuring functions
not listed
At least once every two years, or when you suspect that the
measured value is not accurate.
NIBP leakage test
At least once every two years, or follow hospital regulations
NIBP calibration
At least once every two years, or follow hospital regulations
16.4 ECG calibration
In the process of using the monitor, the displayed ECG signals may be inaccurate due to hardware or software problems, mainly
shown as waveform amplitude becoming larger or smaller. At this moment, you need to calibrate ECG.
Prepare the following instruments before testing:
ECG simulator
ECG cable
Vernier caliper
The calibration method is as follows:
Connect the ECG cable to the monitor.
Connect the ECG electrodes to the ECG simulator.
Select
【
Settings
】
Smart Hotkey
→
“Settings” Menu;
Select “User Maintenance >>”
→
enter the password and confirm
→
“User Maintenance” menu.
Select “Module Maintenance >>”
→
“Module Maintenance” menu.
Select “ECG ”
→
“ECG Maintenance” menu, and select “Calibration” to calibrate the ECG.
Measure the wave amplitude with a caliper; in different filtering modes, ×0.25 is 2.5 ± 5% (mm), ×0.5 is 5.0 ±% 5 (mm), ×1 is
10.0 ±% 5 (mm), and ×2 is 20.0 ±% 5 (mm). Comparing the amplitude of the square wave with the ruler, the error range
should be within 5%.
When calibration is complete, select “Stop Calibration” to exit.
17 Accessories
WARNING
•
Use the accessories specified in this manual. Using other accessories may damage the monitor, or cannot reach
the safety and performance claimed in this manual.
•
The operating and storage environment of the monitor should meet the requirements of the accessories. Please
refer to the manual of the accessories for these requirements.
•
Disposable accessories can only be used once, because repeated use can cause performance degradation.
•
If the packaging or accessories have any sign of damage, do not use such accessories.
•
For ECG Cables, SpO
2
Sensor, Blood Pressure Cuff and Temperature Sensor, the normal life time is two years.
Please replace in time.