Bistos BT-780 Operation Manual Download Page 33

BT-780 Operation manual

 

 

32 

P/N : 780-ENG-OPM-EXP-R00 

Bistos Co., Ltd. 

2020.03 

10.7 Alarm setup 

In “SpO

2

 Setup” menu, select “Alarm Setup >>” to enter “Alarm Setup” interface, and set SpO

2

 alarm switch, alarm level, upper and 

lower alarm limit. See 6.5 Alarm Setup for detailed setting method. 

 

10.8 Technical description 

 

Accessories have passed the biocompatibility test and meet the requirements of ISO 10993-1. 

 

Fluke's index 2XL Oxygen Analyzer can be used to check the function of the monitor and can be used to assess the accuracy of 

the pulse rate but cannot be used to assess the accuracy of blood oxygen. 

 

Functional testers cannot be used to evaluate the accuracy of pulse oximetry probe and pulse oximetry. 

 

Measure the maximum temperature between the oxygen probe and the tissue contact surface

Measured as described in 

Annex BB of ISO 80601-2-61,the temperature is less than 41 ° C. 

 

 

11 NIBP 

11.1 Overview 

The monitor uses oscillometric method to measure noninvasive blood pressure (NIBP). 

The oscillometric method for measuring blood pressure is to inflate a cuff with a certain amount of pressure until the arterial blood 

flow has been completely blocked. As applied pressure decreases, the arterial blood flow which was completely occluded gradually 

opened, and completely opened. Then, the pulsation of the arterial vascular wall will generate a shock wave in the cuff. SBP, MAP, 

and DBP are obtained by measuring and analyzing cuff pressure oscillations when deflating. 

 

Produce first most clear signal - reflect SBP 

 

Oscillation amplitude reaches the peak - reflect MAP 

 

When the cuff pressure is suddenly lowered - reflect DBP 

Measuring mode: manual, cycle, and continuous. Each mode shows systolic, mean and diastolic blood pressure. 

 

Manual mode  

Using Manual mode start to measures by hand 

 

Automatic mode measures  

Use manual mode to open automatic mode, then the measure will automatically turn to automatic mode after a certain time. 

During measurement, any error will stop the current automatic measurement, but not affect next automatic measurement 

unless the time interval less than 30s. If the time interval less than 30s, should delay the next automatic measurement, keep 

the interval more than 30s. 

The time interval can be chosen In Automatic mode as 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 minutes. 

 

Continuous mode  

Choose continuous mode, 5 seconds after complete a measurement start the next measurement, continue 5 minutes then 

stop. During measurement, any error will stop the continuous measurement. If the first measurement time is over 4 minutes 

and 40 seconds but less than 5 minutes, the continuous mode will stop before 5 minutes, if the first measurement time is 

over 5 minutes, the continuous mode will stop after 5 minutes. 

The monitor is defibrillation proof, so the monitor operates normally after defibrillation. 

 

11.2 Safety information

 

  

WARNING

 

 

Do not carry out non-invasive blood pressure measurement on patients with sickle cell disease and skin damage 

or any expected damage. Do not measure NIBP on traumatic body part. This may cause further injury. 

 

When pediatric patients are measured, in order to ensure the cuff pressure does not exceed its maximum 

measurement range of patient types (Adult mode: 300mmHg and Pediatric mode: 240mmHg, Neonate mode: 

150mmHg), you must ensure that you have selected the correct patient type (see patient information menu 

settings). Using the wrong type of pattern is likely to endanger the patient to patient safety, as higher blood 

pressure levels for adults does not apply to pediatric and neonate. 

 

For patients with severe coagulation disorder, determine if the automatic blood pressure measurement is 

carried out according to the clinical evaluation, since the friction of body and cuff may produce hematoma. 

 

Do not install a cuff on the limbs with intravenous infusion or duct, because it may lead to tissue damage around 

the duct when the cuff is inflated and makes the infusion slow down or be blocked. 

 

The inflatable tube connecting the blood pressure cuff and the monitor should be smooth without 

entanglement. The pressure generated by being kinked connection tubing may cause blood flow interference.

 

 

For patients with severe thrombotic disorders, determine whether to carry out automatic blood pressure 

measurement according to the clinical situations, since the limb bundled with a cuff may produce hematoma. 

 

Measure blood pressure frequently will affect the distribution of blood flow, May endanger the safety of 

patients. 

 

Check the patient's physiological condition before measure blood pressure, in order to ensure that long time 

measure will not damage the circulation of patients 

 

For mastectomy patients, applying the NIBP cuff on the surgery side arm can cause lymphedema. Measure blood 

pressure on opposite side arm. 

 

Pressurization of the cuff can temporarily cause loss of function of simultaneously used monitoring device on the 

Summary of Contents for BT-780

Page 1: ...BT 780 Patient monitor Operation Manual Keep this manual for future reference P N 780 ENG OPM EXP R00...

Page 2: ...r no responsibility is assumed by Bistos for its use or any infringements of patents or other rights of third parties that may result from its use No license is granted by implication or otherwise und...

Page 3: ...ng waves 16 3 7 Other common setup 16 4 Patient information management 17 4 1 Patient setup menu 17 4 2 Admitting a patient 17 4 3 Patient information 18 4 4 Discharging a patient 18 4 5 Clear alarms...

Page 4: ...Review 35 13 1 Reviewing trend chart 36 13 2 Reviewing trend table 36 13 3 NIBP measurement review 36 14 Battery 37 14 1 Overview 37 14 2 Battery usage guide 37 14 3 Checking battery performance 37 14...

Page 5: ...nown this information thoroughly before using BT 780 IPX1 Indicates the protection level against the ingress of liquid IPX1 is protection against some water drops falling vertically It correspond the...

Page 6: ...val is once per year If the hospital or healthcare institutions using this device fail to implement a satisfactory maintenance schedule it will result in device failure and may endanger the patient s...

Page 7: ...f the person or organization who makes an unauthorized modification or incorporates an unapproved attachment to the device An operator may only perform maintenance procedures specifically described in...

Page 8: ...mental temperature exceeds 60 Do not heat or splash the battery or throw it into fire or water Do not destroy the battery Do not pierce battery with a sharp object such as a needle Do not hit with a h...

Page 9: ...tory illness e g asthma chronic obstructive pulmonary disease Monitoring during administration of respiratory depressant drugs e g opiate epidural or patient controlled analgesia Assessing oxygen satu...

Page 10: ...ture recordings To enable close observation in resolving hypothermia hyperthermia To observe and monitor patients for changes indicating an infection To monitor the effect of treatment for antimicrobi...

Page 11: ...e1 2 Side view Figure1 3 Rear view 1 5 Description of monitor Figure1 4 Front view Name Description 1 Alarm indicator Indicates the priority of physiological alarm and technical alarms in different co...

Page 12: ...Rear view Name Description 1 Air outlet Heat dissipation 2 Handle Handle for main body transport 3 Speaker holes For alarm and synchronizing sound 4 ID label Identify the monitor information 5 USB po...

Page 13: ...f each parameter module The parameters are shown in fixed position that is from top to bottom and from left to right ECG NIBP SpO2 and PR TEMP RESP 4 Information Tip Area Display the network status ba...

Page 14: ...n any organization or individual shall not interpolate copy or exchange by any means or form When the monitor is combined with other devices it must comply with IEC 60601 1 2005 A1 2012 and should not...

Page 15: ...let If using battery power make sure the battery is fully charged Check all the functions required for patient monitoring to make sure that the monitor operates properly WARNING If the monitor is dama...

Page 16: ...enus is basically similar to the Settings mode as shown below You can set up the monitor screen if you need Menu title A title of the current menu Close menu Close the current menu Exit the current me...

Page 17: ...e screen Select Department Enter the sector and department using the device through the soft keyboard on the screen Select Bed Number Enter the bed number through the soft keyboard on the screen 3 7 2...

Page 18: ...y is very important If the monitor has admitted the patient it is recommended to operate the monitor to discharge the current patient before connecting to not admitted the next patient Otherwise the d...

Page 19: ...onate 7 Select Pace Maker and set whether the patient wears a pacemaker 8 Select Height cm and set the patient s height via the pop up keyboard on the screen Range 0 250 9 Select Weight kg and set the...

Page 20: ...the display format according to needs and get different screen information 5 1 Selecting user interface Select the user interface as follows Select Displays smart hotkey Screen Select Select the displ...

Page 21: ...s the alarm limit or the patient has physiological abnormalities The information of physiological alarm is displayed in the physiological alarm area on top of the screen Technical alarms Technical ala...

Page 22: ...ority alarm Beep beep beep Low priority alarm Beep 6 3 3 Alarm information Alarm information displayed on the physiological or technical alarm area of the monitor indicates the corresponding alarm inf...

Page 23: ...utes 6 4 4 Alarm off As shown in 6 4 3 if the Alarm Pause Time is set to Permanent press the Pause smart hotkey and the monitor will turn off the alarm In this case except the alarm prompt characteris...

Page 24: ...tive range of the system and the upper alarm limit must be higher than the lower alarm limit Select SpO2 Low Limit and set the lower limit of SpO2 alarm Select SpO2 High Limit and set the upper limit...

Page 25: ...for defibrillation proof so the monitor operates normally after defibrillation NOTE In the factory setup ECG wave display in the first two waves from top in the wave area in the normal display format...

Page 26: ...isual alarm 7 3 2 Selecting lead Select the ECG parameter area or wave area ECG Setup menu Select Other Setup ECG Other Setup menu Select Lead Type and select the ECG lead as needed 3 Lead 3 lead ECG...

Page 27: ...art hotkey and select Patient Info menu Select Yes No for Pace Maker as needed indicating that the patient with or without pacemaker Diagnostic Monitor Surgery will not affect rejection of pacemaker p...

Page 28: ...be set Auto 0 25 0 5 1 2 4 Select Wave Speed and set the wave speed The wave speed is 12 5mm s 25mm s 50mm s The default is 25mm s Select Filter and set the filter mode Diagnostic Diagnostic mode Mon...

Page 29: ...ion electrode placement 8 3 1 Adjusting position of respiration electrode If you want to measure ECG and respiration simultaneously you may need to adjust the position of the two electrodes for respir...

Page 30: ...ty In RESP Setup menu select sensitivity and set the sensitivity The sensitivity is 1 2 3 4 5 The default is 2 8 6 Alarm setup Select Alarm Setup Alarm Setup interface to set respiration related alarm...

Page 31: ...will have deviation When the patient has a tendency to hypoxia use the oximeter to analyze blood samples in order to fully grasp the patient s condition Do not put the sensor on limbs with arterial du...

Page 32: ...ncy radio wave interference such as interference generated by the host system or interference from electrosurgery instrument connected to the system Intravenous dye Too frequent movement of the patien...

Page 33: ...fter complete a measurement start the next measurement continue 5 minutes then stop During measurement any error will stop the continuous measurement If the first measurement time is over 4 minutes an...

Page 34: ...beats min the blood pressure measurement is impossible Obese patients A thick layer of fat around a limb blocks the arterial oscillation so that it cannot reach the cuff The accuracy is lower than no...

Page 35: ...tient is neonate the pressure can be select from 100 120 The default cuff pressure value is 100 11 7 NIBP reset Select NIBP parameter area NIBP Setup menu Select Reset and restore the inflation pressu...

Page 36: ...mperature in the specified environment temperature range or else it may be dangerous 12 5 Temperature display The monitor can display the body temperature of two channels T1 and T2 and the alarm limit...

Page 37: ...r values of the time which will change automatically with the move of trend cursor 13 2 Reviewing trend table Select Trend smart hotkey to enter Trend menu select Tabular and enter the following windo...

Page 38: ...ry out of the reach of children Use only the designated battery If the battery is damaged or leaks replace it immediately Do not use a defective battery for the monitor Do not open the battery compart...

Page 39: ...manufacturer s instructions Do not use abrasive materials such as steel wool Do not allow any liquid entering into the enclosure and never immerse any part of the device into liquid Do not leave any c...

Page 40: ...ue is not accurate NIBP leakage test At least once every two years or follow hospital regulations NIBP calibration At least once every two years or follow hospital regulations 16 4 ECG calibration In...

Page 41: ...In accordance with classification specified in the European Medical Device Directive 93 42 EEC this monitor is Class IIb device The monitor is classified as follows in accordance with IEC 60601 1 Cate...

Page 42: ...mm mV x0 5 10mm mV x1 0 20mm mV x2 0 40mm mV x4 0 Wave sweep speed 12 5 mm s 25 mm s 50 mm s Band width Diagnostic mode 0 05 100 Hz Monitor mode 0 5 40 Hz Surgery mode 1 25 Hz CMRR 100 dB Notch 50 60...

Page 43: ...Resolution 0 1 Measurement accuracy 0 3 Number of channel 2 Alarm T1 T2 upper limit 0 1 50 0 0 1 step T1 T2 lower limit 0 49 9 0 1 step TD upper limit 0 50 0 0 1 step RESP Measurement method Thoracic...

Page 44: ...ring with arterial blood sample reference measured with a CO oximeter Pulse oximeter measurements are statistically distributed and about two thirds of the measurements are expected to come within the...

Page 45: ...orks abnormally check the pacemaker Pacemaker Not work H ECG Signal weak H The patient ECG signal is too weak and the system can t analyze Check the patient s condition electrodes cables and leads ST...

Page 46: ...s Manual is used Check the sensor mounting position whether the sensor is damaged or sensor type Reconnect the sensor or use new sensor SpO2 Sensor Off L SpO2 Search Timeout L SpO2 Search Pulse L Sens...

Page 47: ...settings On On On On Mid Mid Mid Mid Off Off Off Off Gain x1 x1 x1 Wave velocity 25 0mm s 25 0mm s 25 0mm s Filter mode Monitor Monitor Monitor Wave color Green Green Green Wave style Color scale Col...

Page 48: ...whether the pressure sensor is normal If the above doesn t solve the problem please contact Bistos after sales department or dealers 22 Manufacturer s declaration on EMC BT 780 needs special precauti...

Page 49: ...and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection...

Page 50: ...is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transm...

Page 51: ...g BT 780 This product is manufactured and passed through strict quality control and inspection Compensation standard concerning repair replacement refund of the product complies with Framework Act on...

Reviews: