BT-780 Operation manual
17
P/N : 780-ENG-OPM-EXP-R00
Bistos Co., Ltd.
2020.03
3.
Pulse Volume
Select
【
Volume
】
smart hotkey
→
“Volume Setup” menu.
Select “Pulse Volume”: Set pulse volume from 0 to 9. 0 means off.
4.
Touch Volume
Select
【
Volume
】
smart hotkey
→
“Volume Setup” menu.
Select “Touch Volume”: Set touch volume from 0 to 9. 0 means off.
5.
Key Volume
Select
【
Volume
】
smart hotkey
→
“Volume Setup” menu.
Select “Key Volume”: Set key volume from 0 to 9. 0 means off.
3.7.5 Setting parameter unit
You can select a preferred unit through the following operations
Select
[Settings]Smart Hotkey
→
“
Settings
”
.
Select “User Maintenance >>”
→
enter the password and confirm
→“
User Maintenance” menu.
Select “Unit Setup >>”
→“
Unit Setup” menu.
Select “Height Unit”, and select the unit “cm” / “inch” as needed.
Select “Weight Unit”, and select the unit “kg” / “lb” as needed.
“ST Unit” fixed as “mV”, is not optional.
Select “Pressure Unit”, and select the unit “mmHg” / “kPa” as needed.
Select “TEMP Unit”, and select the unit “
℃
” / “
℉
”as needed.
4 Patient information management
Connect the patient to the monitor, and the monitor will display and store the physiological data of the patient, so the patient can
be monitored without admitting the patient. However, admitting the patient correctly is very important.
If the monitor has admitted the patient, it is recommended to operate the monitor to discharge the current patient before
connecting to (not admitted) the next patient. Otherwise, the data of the previous patient will be stored in the data of the current
patient.
WARNING
Whether the patient is admitted or not, the system will give a default value to “Patient Type” and “Pace Maker”,
“Patient Type” default “Adult”, “Pace Maker” default “No”, and the user must confirm that the default value is
appropriate for the patient being monitored.
For patients with pacemakers, “Pace Maker” must be set to “Yes”. Otherwise, the pacing pulse will be treated as
normal QRS wave group, and the system is unable to detect the alarm status of “ECG Signal weak”.
For patients without a pacemaker, “Pace Maker” must be set to “No”. Otherwise, the system is unable to detect
the arrhythmias (including PVCs count) related to ventricular premature beats, and fails to perform ST segment
analysis.
4.1 Patient setup menu
You can manage the patient through the
“
Patient” menu. To enter
“
Patient” menu, operate as follows:
Select
[Settings]Smart Hotkey
→
“
Settings
”
→
“
Patient >>
”
→
“
Patient
”
menu;
Or Select
[Pat. Set] Smart Hotkey to enter “Patient” menu, as shown in Fig. 4-1.
Figure 4-1 “Patient” menu
4.2 Admitting a patient
Admit a patient as follows:
In “Patient” menu, select “Quick Admit”
→
“Warning” message
→
“OK”
→
“Quick Admit” menu, as shown in Figure 4-2.