background image

BT-780 Operation manual

 

 

30 

P/N : 780-ENG-OPM-EXP-R00 

Bistos Co., Ltd. 

2020.03 

10 SpO

2

 

10.1 Overview 

Blood oxygen saturation (SpO

2

) is the percentage of oxyhemoglobin (HbO2) capacity bound by oxygen in the blood in the total 

hemoglobin (Hb) capacity that can be combined, that is, the concentration of oxygen in the blood. 

The principle for monitoring the pulse SpO

2

 is to fix the probe fingerstall on the patient’s finger or toe, use the finger (or toe) as a 

transparent container for hemoglobin, use 660nm wavelength red light and 950nm near-infrared light as the incident light, 

maximum output power is 300 mW, measure the light transmission intensity through the tissue bed, and calculate the 

concentration of hemoglobin and SpO

2

The passing lights depend on a variety of factors, most of which are constant. However, one of these factors, the arterial blood flow, 

changes with time, as it is pulsating. By measuring the light absorbed during pulsating, it is possible to obtain the arterial blood 

SpO

2

. Detection pulsation can give a “plethysmography” wave and pulse rate signal. 

The main screen displays “SpO

2

” value and “plethysmography” wave. 

This monitor applies to measure SpO

2

 of adults (>18 years) and pediatric (<18 years,>30 days), neonate (<30 days). Contact SpO

2

 

probe to Patient’s finger (or toe) to get “SpO

2

” value and “plethysmography” wave. 

SpO

2

 function of this monitor has been calibrated in factory. 

The monitor is defibrillation proof, so the monitor operates normally after defibrillation. 

 

10.2 Safety information

 

  

WARNING

 

 

Please use SpO2 sensor specified in this Manual, operate in accordance with the Manual, and observe all 

warnings and precautions. 

 

Before monitoring, check whether the sensor cable is normal. When SpO2 sensor cable is unplugged from the 

socket, the screen will display 

SpO2 Sensor Off” error message, and trigger an audible and visual alarm 

simultaneously. 

 

If the sensor or sensor packaging has signs of damage, do not use this SpO2 sensor; return it to the 

manufacturer. 

 

If there is carboxyhemoglobin, methemoglobin or dye diluted chemical, the SpO2 value will have deviation. 

 

When the patient has a tendency to hypoxia, use the oximeter to analyze blood samples in order to fully grasp 

the patient's condition. 

 

Do not put the sensor on limbs with arterial duct or intravenous tube. 

 

Do not intertwine electrosurgical equipment cable with the sensor cable. 

 

Avoid using the monitor and sensors while using the NMR equipment, in order to avoid severe burns to the 

patient as a result of induced currents. 

 

During long time continuous monitoring of a patient, check the position of SpO2 sensor once every 2 hours, and 

move properly when the skin changes or every four hours. Some patients may require more frequent inspection, 

such as patients with perfusion disorders or sensitive skin, because persistent and prolonged monitoring may 

increase unpredictable skin changes, such as allergies, redness, blistering or pressure necrosis. 

 

When the measured pulse rate is not complete, then the "---". 

 

Before using, verify compatibility between the monitor, probe and cable, otherwise it may cause injury to the 

patient. 

 

Functional testers cannot be used to evaluate the accuracy of pulse oximetry and pulse oximetry. 

 

SpO2 low alarm limit cannot be less than 85. 

 

NOTE

 

 

Do not put the oxygen probe and blood pressure cuff on the same limb, because blood flow occlusion during 

blood pressure measurement will affect the SpO

2

 readings. 

 

The monitor cannot be used to verify the accuracy of SpO

2

 probe and SpO

2

 equipment. 

 

10.3 Monitoring steps 

1. 

Select the appropriate SpO

2

 sensor according to the patient. 

2. 

Turn on the monitor, and connect the SpO

2

 lead wire to the monitor. 

3. 

Clean the measurement site, such as finger with nail polish. 

4. 

Put the SpO

2

 sensor probe on the patient’s body. 

5. 

Select the appropriate alarm settings. 

6. 

Start monitoring. 

NOTE

 

 

Turn on the monitor, plug in SpO

2

 probe and connect patient’s finger (or toe), monitor displays SpO

wave, “SpO

2

 

Pulse Search” displayed in the technical alarm area until the monitor measured SpO

2

 value and pulse rate. “SpO

2

 

Search Timeout” displayed in the technical alarm area until the monitor measured pulse rate. Check the sensor 

mounting position, whether the sensor is damaged or sensor type. Reconnect the sensor or use new sensor. 

 

10.4 Display 

SpO2 parameter area is as shown in figure 10-1. 

Summary of Contents for BT-780

Page 1: ...BT 780 Patient monitor Operation Manual Keep this manual for future reference P N 780 ENG OPM EXP R00...

Page 2: ...r no responsibility is assumed by Bistos for its use or any infringements of patents or other rights of third parties that may result from its use No license is granted by implication or otherwise und...

Page 3: ...ng waves 16 3 7 Other common setup 16 4 Patient information management 17 4 1 Patient setup menu 17 4 2 Admitting a patient 17 4 3 Patient information 18 4 4 Discharging a patient 18 4 5 Clear alarms...

Page 4: ...Review 35 13 1 Reviewing trend chart 36 13 2 Reviewing trend table 36 13 3 NIBP measurement review 36 14 Battery 37 14 1 Overview 37 14 2 Battery usage guide 37 14 3 Checking battery performance 37 14...

Page 5: ...nown this information thoroughly before using BT 780 IPX1 Indicates the protection level against the ingress of liquid IPX1 is protection against some water drops falling vertically It correspond the...

Page 6: ...val is once per year If the hospital or healthcare institutions using this device fail to implement a satisfactory maintenance schedule it will result in device failure and may endanger the patient s...

Page 7: ...f the person or organization who makes an unauthorized modification or incorporates an unapproved attachment to the device An operator may only perform maintenance procedures specifically described in...

Page 8: ...mental temperature exceeds 60 Do not heat or splash the battery or throw it into fire or water Do not destroy the battery Do not pierce battery with a sharp object such as a needle Do not hit with a h...

Page 9: ...tory illness e g asthma chronic obstructive pulmonary disease Monitoring during administration of respiratory depressant drugs e g opiate epidural or patient controlled analgesia Assessing oxygen satu...

Page 10: ...ture recordings To enable close observation in resolving hypothermia hyperthermia To observe and monitor patients for changes indicating an infection To monitor the effect of treatment for antimicrobi...

Page 11: ...e1 2 Side view Figure1 3 Rear view 1 5 Description of monitor Figure1 4 Front view Name Description 1 Alarm indicator Indicates the priority of physiological alarm and technical alarms in different co...

Page 12: ...Rear view Name Description 1 Air outlet Heat dissipation 2 Handle Handle for main body transport 3 Speaker holes For alarm and synchronizing sound 4 ID label Identify the monitor information 5 USB po...

Page 13: ...f each parameter module The parameters are shown in fixed position that is from top to bottom and from left to right ECG NIBP SpO2 and PR TEMP RESP 4 Information Tip Area Display the network status ba...

Page 14: ...n any organization or individual shall not interpolate copy or exchange by any means or form When the monitor is combined with other devices it must comply with IEC 60601 1 2005 A1 2012 and should not...

Page 15: ...let If using battery power make sure the battery is fully charged Check all the functions required for patient monitoring to make sure that the monitor operates properly WARNING If the monitor is dama...

Page 16: ...enus is basically similar to the Settings mode as shown below You can set up the monitor screen if you need Menu title A title of the current menu Close menu Close the current menu Exit the current me...

Page 17: ...e screen Select Department Enter the sector and department using the device through the soft keyboard on the screen Select Bed Number Enter the bed number through the soft keyboard on the screen 3 7 2...

Page 18: ...y is very important If the monitor has admitted the patient it is recommended to operate the monitor to discharge the current patient before connecting to not admitted the next patient Otherwise the d...

Page 19: ...onate 7 Select Pace Maker and set whether the patient wears a pacemaker 8 Select Height cm and set the patient s height via the pop up keyboard on the screen Range 0 250 9 Select Weight kg and set the...

Page 20: ...the display format according to needs and get different screen information 5 1 Selecting user interface Select the user interface as follows Select Displays smart hotkey Screen Select Select the displ...

Page 21: ...s the alarm limit or the patient has physiological abnormalities The information of physiological alarm is displayed in the physiological alarm area on top of the screen Technical alarms Technical ala...

Page 22: ...ority alarm Beep beep beep Low priority alarm Beep 6 3 3 Alarm information Alarm information displayed on the physiological or technical alarm area of the monitor indicates the corresponding alarm inf...

Page 23: ...utes 6 4 4 Alarm off As shown in 6 4 3 if the Alarm Pause Time is set to Permanent press the Pause smart hotkey and the monitor will turn off the alarm In this case except the alarm prompt characteris...

Page 24: ...tive range of the system and the upper alarm limit must be higher than the lower alarm limit Select SpO2 Low Limit and set the lower limit of SpO2 alarm Select SpO2 High Limit and set the upper limit...

Page 25: ...for defibrillation proof so the monitor operates normally after defibrillation NOTE In the factory setup ECG wave display in the first two waves from top in the wave area in the normal display format...

Page 26: ...isual alarm 7 3 2 Selecting lead Select the ECG parameter area or wave area ECG Setup menu Select Other Setup ECG Other Setup menu Select Lead Type and select the ECG lead as needed 3 Lead 3 lead ECG...

Page 27: ...art hotkey and select Patient Info menu Select Yes No for Pace Maker as needed indicating that the patient with or without pacemaker Diagnostic Monitor Surgery will not affect rejection of pacemaker p...

Page 28: ...be set Auto 0 25 0 5 1 2 4 Select Wave Speed and set the wave speed The wave speed is 12 5mm s 25mm s 50mm s The default is 25mm s Select Filter and set the filter mode Diagnostic Diagnostic mode Mon...

Page 29: ...ion electrode placement 8 3 1 Adjusting position of respiration electrode If you want to measure ECG and respiration simultaneously you may need to adjust the position of the two electrodes for respir...

Page 30: ...ty In RESP Setup menu select sensitivity and set the sensitivity The sensitivity is 1 2 3 4 5 The default is 2 8 6 Alarm setup Select Alarm Setup Alarm Setup interface to set respiration related alarm...

Page 31: ...will have deviation When the patient has a tendency to hypoxia use the oximeter to analyze blood samples in order to fully grasp the patient s condition Do not put the sensor on limbs with arterial du...

Page 32: ...ncy radio wave interference such as interference generated by the host system or interference from electrosurgery instrument connected to the system Intravenous dye Too frequent movement of the patien...

Page 33: ...fter complete a measurement start the next measurement continue 5 minutes then stop During measurement any error will stop the continuous measurement If the first measurement time is over 4 minutes an...

Page 34: ...beats min the blood pressure measurement is impossible Obese patients A thick layer of fat around a limb blocks the arterial oscillation so that it cannot reach the cuff The accuracy is lower than no...

Page 35: ...tient is neonate the pressure can be select from 100 120 The default cuff pressure value is 100 11 7 NIBP reset Select NIBP parameter area NIBP Setup menu Select Reset and restore the inflation pressu...

Page 36: ...mperature in the specified environment temperature range or else it may be dangerous 12 5 Temperature display The monitor can display the body temperature of two channels T1 and T2 and the alarm limit...

Page 37: ...r values of the time which will change automatically with the move of trend cursor 13 2 Reviewing trend table Select Trend smart hotkey to enter Trend menu select Tabular and enter the following windo...

Page 38: ...ry out of the reach of children Use only the designated battery If the battery is damaged or leaks replace it immediately Do not use a defective battery for the monitor Do not open the battery compart...

Page 39: ...manufacturer s instructions Do not use abrasive materials such as steel wool Do not allow any liquid entering into the enclosure and never immerse any part of the device into liquid Do not leave any c...

Page 40: ...ue is not accurate NIBP leakage test At least once every two years or follow hospital regulations NIBP calibration At least once every two years or follow hospital regulations 16 4 ECG calibration In...

Page 41: ...In accordance with classification specified in the European Medical Device Directive 93 42 EEC this monitor is Class IIb device The monitor is classified as follows in accordance with IEC 60601 1 Cate...

Page 42: ...mm mV x0 5 10mm mV x1 0 20mm mV x2 0 40mm mV x4 0 Wave sweep speed 12 5 mm s 25 mm s 50 mm s Band width Diagnostic mode 0 05 100 Hz Monitor mode 0 5 40 Hz Surgery mode 1 25 Hz CMRR 100 dB Notch 50 60...

Page 43: ...Resolution 0 1 Measurement accuracy 0 3 Number of channel 2 Alarm T1 T2 upper limit 0 1 50 0 0 1 step T1 T2 lower limit 0 49 9 0 1 step TD upper limit 0 50 0 0 1 step RESP Measurement method Thoracic...

Page 44: ...ring with arterial blood sample reference measured with a CO oximeter Pulse oximeter measurements are statistically distributed and about two thirds of the measurements are expected to come within the...

Page 45: ...orks abnormally check the pacemaker Pacemaker Not work H ECG Signal weak H The patient ECG signal is too weak and the system can t analyze Check the patient s condition electrodes cables and leads ST...

Page 46: ...s Manual is used Check the sensor mounting position whether the sensor is damaged or sensor type Reconnect the sensor or use new sensor SpO2 Sensor Off L SpO2 Search Timeout L SpO2 Search Pulse L Sens...

Page 47: ...settings On On On On Mid Mid Mid Mid Off Off Off Off Gain x1 x1 x1 Wave velocity 25 0mm s 25 0mm s 25 0mm s Filter mode Monitor Monitor Monitor Wave color Green Green Green Wave style Color scale Col...

Page 48: ...whether the pressure sensor is normal If the above doesn t solve the problem please contact Bistos after sales department or dealers 22 Manufacturer s declaration on EMC BT 780 needs special precauti...

Page 49: ...and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection...

Page 50: ...is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transm...

Page 51: ...g BT 780 This product is manufactured and passed through strict quality control and inspection Compensation standard concerning repair replacement refund of the product complies with Framework Act on...

Reviews: