TX-KFP
Bionics Instrument
© 2004 Eur 20040426, all rights reserved
- 23 -
Fig. 3.4.1
Pin 6
Pin 5
Black
Brown
White
(Shield)
Cable by client
From KFP unit
1 2 3
1 2 3
Fig. 3.4.2
3. Installation
3.4 Electrical Connections General
- Cabling:
The TX-KFP is supplied with a power/signal cable of approximately 2.5 mtr. long.
One end is fitted with a DIN connector, the other end is open.
Cable
connection
specification:
White
-
+
24
Volt
Brown
-
4
–
20
mA
Black - Cable shield
SV-60 DIN connector specifications:
Pin
1-4: spare
Pin
5:
+24
Volt
Pin
6:
4
–
20
mA
output
- Junction Box:
The standard TX-KFP product is supplied with a junction box.
The 2.5 m 2-wire/ shielded (power/signal) cable must be connected to the junction box
terminals as specified below.
Terminal Specification:
1 - 24 VDC in/out (White)
2 - 4-20 mA in/out (Brown)
3 – Shield (Black)