3. Installation
3.3 Duct mounting of the TX-KFP in exhaust systems
The installation of a TX-KFP in an exhaust system requires a 40 mm diameter stub provided
with a KF-40 type flange installed on the pipe- or duct-work in order to match the KF-40
type body of the TX-KFP gas detector.
`
Duct mounting details:
Place the gasket (3) in the groove of the KF-40 type
flange of the stub and then insert the TX-KFP unit.
After insertion of the TX-KFP unit into the stub secure the
two flange parts with the clamping ring (see fig. 3.3.2).
1 – TX-KFP
2 – Clamping ring
3 – Gasket (Centring ring with O- ring)
4 – KF-40 stub
5 – Exhaust duct
Fig. 3.3.1
Fig. 3.3.2
To ensure that the sensor is sufficiently in contact with
the flow of air/gas, the flange should have maximum
depth “L
max
“ as illustrated (see fig. 3.3.1)
“L
max
”
= Sensor length – 5 mm