3. Installation
3.4.1 Wiring details for PC-5028 board
The PC-5028 board is part of the wall/ceiling mount adapter (see fig. 3.4.3)
3.4.2 Wiring details for analogue/digital local indicator
For local readings of the detected gas level or for calibration purposes an analogue or digital
local indicator may be installed.
In- and out-going cables must be connected to the top and bottom terminals as specified
below (see fig. 3.4.4)
Wiring details for connecting
the controller/signal cable to
connector CN2 .
Terminal specification:
1: not used
2: not used
3: 4 -20 mA
4: 24V DC
5: not used
Fig. 3.4.3
Fig. 3.4.4
Terminal specifications:
Top (TX-KFP cable):
1: 24V DC (White)
2: mA out )((Brown)
3: in use for indicator
4: shield (Black)
Bottom (Client’s cable)
1: 24 V DC
2: in use for indicator
3: mA
4: shield