3. Installation
3.2 Mounting the gas detection unit
The TX-KFP unit can be installed as an environmental gas detector, whereby a choice can be
made between wall- or ceiling– mounting the unit or the gas detector may be duct-mounted
for monitoring applications in exhaust systems.
Fig. 3.2.1
The TX-KFP unit should preferably
be installed vertically.
In case the unit needs to be installed
under an angle, please make sure
that the unit is never placed under
an angle greater than 45° relative to
a vertical position (see figure 3.2.1).
This is to ensure that the main
electrode is always in contact with a
sufficient amount of electrolyte.
If the TX-KFP is placed under an
angle the ’pressure stabilization
screw ‘ should always face up
wards.